7 Disassembly
Removing the ventilator and MEDUCORE Standard
Applies to
•
MEDUMAT Standard
2
•
MEDUMAT Easy
CPR
•
MEDUCORE Standard
Required auxiliary
equipment
Screwdriver with crosshead PH1
Requirement
Disassembly is prepared
(see "7.5.1 Preparing disassembly",
.
1. With LIFE-BASE 1 NG XL:
•
Undo the knurled screws.
•
Pull out the hangers.
•
Tighten the knurled screws.
2. Place the portable unit and devices with the control panel
membrane facing down.
Figure without devices