10 Alarms and error messages
Flow measurement is not
functioning
Flow measu ASB option
is deactivated in the operator
menu
Activate the Flow
measu ASB option in the
operator menu
FlowCheck sensor connection line/
FlowCheck sensor connection line
with MEDUtrigger defective
Replace the FlowCheck sensor
connection line/FlowCheck sensor
connection line with MEDUtrigger
CO
2
measurement is not
functioning
Capnography option is deactivated
in the operator menu
Activate the MEDUtrigger option
in the operator menu
.
Esophagus intubation
Check correct intubation.
CO
2
measuring tube not
connected correctly
Check CO
2
measuring tube.
Ongoing occlusion
Eliminate the occlusion.
No ventilation in CCSV
Endotracheal tube not sufficiently
blocked
Check cuff pressure.
Esophagus intubation
Check correct intubation.
Chest compression is not being
performed
Restart chest compression.
Use breathing circuit with reduced
dead space or the 3 m breathing
circuit under CCSV
•
Use standard 2 m breathing
circuit.
•
Switch to IPPV mode.
Fault
Cause
Remedy