14 Appendix
Monitoring
Pressure gauge:
Airway pressure
Frequency tachometer (only with CCSV option)
Curves:
•
Airway pressure (only with Flow measu ASB option
and Curve display option or Capnography option)
•
Flow (only with Flow measu ASB option and Curve
display option)
•
CO
2
(only with Capnography option)
•
etCO
2
trend (only with Capnography option)
Monitoring parameters
pPeak, pPlat, pMean
Vte, MVe, f, fsp, Vleak (only with Flow measu ASB
option)
etCO
2
(only with Capnography option)
Operating gas
Medical oxygen (100 % oxygen)
or
Concentrator oxygen as per the European Pharmacopeia
(90 % to 96 % oxygen)
Operating pressure range
2.7 bar to 6 bar
Recommended gas supply
4.5 bar at
≥
150 l/min
Required gas supply
min. 2.7 bar at
≥
80 l/min
Maximum outlet flow
80 l/min with inlet pressure 4.5 bar in Air Mix and No Air Mix
operation
Mechanical pressure relief/emergency
valve
Pressure limitation to a maximum of 100 mbar
I:E
1:1 in CPR 30:2, 15:2 and RSI Manual
1:4 to 4:1*
* The setting options of I:E in volume-controlled ventilation modes
depend on the combination of ventilation rate and tidal volume.
Ventilation rate
5 min
-1
to 50 min
-1
(±1 min
-1
)
Inspiration time
0.2 s to 9.6 s
Tidal volume
(1)
50 ml to 2000 ml (±40 ml or ±20 %)
Respiratory minute volume
(1)
At least 0.25 l
Max. 20 l
Gas composition
Air, oxygen, CO
2
mix. Oxygen percentage 21 % to 100 %,
CO
2
percentage 0 % to 10 %
Inspiration pressure (pInsp) (only with
Pressure controlled ventilation modes
option)
3 mbar to 60 mbar (±3 mbar or ±15 %)
Maximum ventilation pressure (pMax) 10 mbar to 65 mbar (±3 mbar or ±15 %)
Minimum working pressure
3 mbar. Cannot be set.
Specification
Device