260
EN MEDUMAT
Standard
2
WM
6807
1
0
5
/2
021
14 Appendix
14.3.2 Recorded mission logs
Mission logs contain detailed session data from up to 100
applications. The number of saved applications may vary
depending on the session duration.
Depending on the frequency of the applications, the time required
to export the data may vary.
The following data are saved in the mission logs:
•
Measurements: The device records average values of the
measured values as trend data.
•
Ventilation settings and their changes: All triggered alarms and
settings changes are saved immediately.
•
Triggered alarms
alarmsystem
Test of the visual and audible alarms
buttontest
Test of the buttons and navigation knobs
temperature sensor
Test of the internal temperature of the device
airway / mixing chamber
pressure sensors
Test of the internal pressure sensors
int./ext. flow sensor
Test of the internal flow sensor
pressure drop
Test of the pneumatic bleed time
leak tightness
Test of the tightness of the device including the breathing circuit
patienthose obstructed
Test of the patency of the breathing circuit
input pressure sensor
Test of the inlet pressure sensor
airmix valve
Test of the Air Mix operation
flowcheck sensor
Test of the FlowCheck sensor
flowcheck cable
Test of the FlowCheck sensor connection line/FlowCheck sensor connection
line with MEDUtrigger
flowcheck offset
Test of the offset of the FlowCheck sensor
flowcheck sensor sn
Documentation of the serial number of the FlowCheck sensor which was
connected during the function check
co2system
Test of the CO
2
module
volume output too low
Test of the volume output
Column name
Description
Summary of Contents for MEDUMAT Standard2
Page 1: ...MEDUMAT Standard2 Ventilator Instructions for Use for Devices from Software Version 5 1...
Page 269: ......
Page 270: ......
Page 271: ......