4 Preparation and operation
6. With the reusable breathing circuit: Guide the CO
2
measuring
tube with the rest of the measuring circuit and ventilation hose
into the protective sleeve of the breathing circuit.
7. If necessary: Activate Capnography option
(see "6.3.9 Options", page 138)
Result
The CO
2
measuring tube is connected to the breathing circuit.
4.4.5 Connecting MEDUtrigger
The operational readiness of MEDUtrigger is indicated by 2 green
LEDs on MEDUtrigger. If MEDUtrigger is connected to the device
and the green LEDs on MEDUtrigger are lit, you can trigger
mechanical breaths manually by pressing the MEDUtrigger button.
4-2 Connection line of MEDUtrigger (
) and FlowCheck sensor
connection line with MEDUtrigger (
To minimize the dead space, you can also connect the CO
2
measuring tube to a breathing system filter with gas connection
(e.g., WM 22162).
MEDUtrigger button
LEDs
1
2