4 Preparation and operation
4.4.8 Connecting the nebulizer
Only use the device in combination with the following nebulizer:
•
Pneumatic medication nebulizer WM 15827
•
Aerogen
®
Solo (Aerogen Ltd.)
•
Tube Inhaler (VBM Medizintechnik GmbH)
1. Place the mask/endotracheal tube
and/or MEDUtrigger
) on the nebulizer
,
or
.
Risk of injury due to erroneous readings!
If the filter is installed incorrectly or no filter is used, the membrane
may stick in the patient valve or the FlowCheck sensor could
return erroneous readings, which can cause injury to the patient.
Observe the correct order of the individual components.
Install the filter (breathing system filter, bacteria filter or a
combined breathing system/bacteria filter) between the
FlowCheck sensor and nebulizer.
Risk of injury from use of pneumatic nebulizers during
volume-controlled ventilation!
The use of pneumatic nebulizers increases the minute volume
administered to the patient. This can injure the patient.
Do not use pneumatic nebulizers during volume-controlled
ventilation.