4 Preparation and operation
4.7.9 Performing ventilation in CPR Manual mode
In CPR Manual mode, you determine the respiratory rate
administered yourself. Using MEDUtrigger, you manually trigger
individual mechanical breaths with the set tidal volume.
Requirement
•
The device is switched on.
•
MEDUtrigger is connected to the device.
•
MEDUtrigger is activated in the operator menu
(Operator menu | Options | MEDUtrigger).
•
Manual
as set in the operator menu (Operator menu |
Presets patient | CPR mode | Start mode).
1. Briefly press the CPR button
.
The device switches to
CPR Man.
mode.
The green LEDs
on MEDUtrigger light up.
If you have activated the metronome, the metronome emits
signals at a rate of 30:2 (patient group Adult) or 15:2
(patient group Infant and Child).
2. Press and hold the MEDUtrigger button during the ventilation
pause until two mechanical breaths are performed.
or
If the green LEDs on MEDUtrigger are lit, briefly press the
MEDUtrigger button twice and trigger the mechanical
breaths manually.
3. If necessary: Cancel the CPR Manual mode with the
CPR button
.
Delay in treatment due to simultaneous metronome outputs
from the ventilator and the defibrillator!
If the ventilator is used together with a defibrillator
(MEDUCORE Standard or MEDUCORE Standard
2
) which can also
emit a metronome sound, the simultaneous metronome outputs
from the defibrillator and the ventilator may confuse the user and
cause delays in treatment.
If the ventilator and defibrillator are being used at
the same time, switch off the metronome sound of
MEDUMAT Standard
2
.
Do not perform resuscitation without a metronome.