Repairs to OXYWAY Fast I, OXYWAY Fast I+, OXYWAY Fast III
14.21 Replacing the threaded pin in the
multistep flow selector
Required tools
14.21.1
Removing the threaded pin from the multistep
flow selector
Requirement
• The multistep flow selector assembly
has been removed
(see "14.13.1 Removing the multistep flow selector assembly",
page 98)
• The spring for the piston
(see "14.14.1 Removing the spring for the piston", page 99)
.
• The piston
has been removed
• The multistep flow selector assembly
has been disassembled
(see "14.17.1 Disassembling the multistep flow selector assembly",
page 101)
1. Use the Allen key to remove the threaded pin
from the
multistep flow selector
.
14.21.2
Fitting the threaded pin in the multistep flow
selector
1. Insert the threaded pin
into the multistep flow selector
and screw in using the Allen key.
When doing so, note: The threaded pin
must not be screwed
into the hole for the orifice plate assembly
2. Assemble the multistep flow selector assembly
(see "14.17.2 Assembling the multistep flow selector assembly",
page 102)
3. Fit the piston
(see "14.16.2 Fitting the piston", page 101)
4. Fit the spring for the piston
(see "14.14.2 Fitting the spring for the piston", page 99)
5. Fit the multistep flow selector assembly
(see "14.13.2 Fitting the multistep flow selector assembly",
page 98)