Repairs to OXYWAY Fast I, OXYWAY Fast I+, OXYWAY Fast III
14.27.1
Disassembling the cover assembly
Requirement
• The multistep flow selector assembly
has been removed
(see "14.13.1 Removing the multistep flow selector assembly",
page 98)
• The spring for the piston
has been removed
(see "14.14.1 Removing the spring for the piston", page 99)
• The piston
has been removed
• The multistep flow selector assembly
(see "14.17.1 Disassembling the multistep flow selector assembly",
page 101)
.
1. Lightly tension the cover clamping device
in the vise.
2. Place the cover assembly
in the cover clamping device
3. Align the cover assembly
in the cover clamping device
so that the cover clamping device
clamps onto the housing
cover
4. Tighten the cover clamping device
in the vise.
5. Screw the special counter tool DISS
outlet
.
6. Position the special gripping pliers
on the union nut
.
7. Use the special gripping pliers
and the special counter tool
DISS
to slightly loosen the union nut
cover
When doing so, note: Do not yet remove the union nut
from the housing cover
8. Remove the special counter tool DISS
from the inhalation
outlet
.
9. Use the open-end wrench to remove the inhalation outlet
from the multistep flow selector
10.Remove the housing cover
from the cover clamping
device
together with the union nut
and the multistep
flow selector
.
11.Unscrew the union nut
from the housing cover
by hand.
12.Remove the union nut
.
13.Remove the multistep flow selector