WM 66
611 07/
202
1
OXYWAY
EN
125
Repairs to OXYWAY Fast II
15 Repairs to OXYWAY Fast II
15.1 General instructions
• When carrying out repairs, please refer to the instructions for use
and the service manual. In particular, you should observe the safety
information contained in the instructions for use.
• Always ensure that the workstation is kept clean and tidy.
• Only carry out the repairs described in this service manual.
• Always carry out a final inspection after each repair job
(see "9 Final inspection OXYWAY Fast II", page 38)
.
• Only use original spare parts from WEINMANN Emergency.
• You will find the necessary auxiliary equipment in the
"Required auxiliary equipment" chapter
(see "21 Required auxiliary equipment", page 251)
• The item numbers in this chapter are identical to the item numbers
in the "Spare parts" chapter
.
• Use a container with compartments for the screws and washers.
• Place all the necessary tools ready before starting work.
15.2 Sequence of repair steps
The flow charts in this chapter provide an overview of the repair steps
that are necessary in order to replace a particular part (shown in gray).
Back
Forward
Previous page
Next page
Summary of Contents for OXYWAY Click
Page 271: ......