Repairs to OXYWAY Fast II
15.15.1
Removing the O
2
coupling
Requirement
The O-ring
has been removed from the connecting bolt
(see "15.8.1 Removing the O-ring from the connecting bolt",
page 134)
1. Clamp the bottom part of the counter tool
in the vise.
2. Put the top part of the counter tool
onto the bottom part of
the counter tool
.
3. Screw the hand-screwed connection
of the pressure reducer
onto the bottom part of the counter tool
by hand.
4. Tightly screw the hand-screwed connection
onto the top part
of the counter tool
.
5. Use the open-end wrench to loosen the O
2
coupling
and
reducer from the housing.
6. Remove the O
2
coupling
and reducer from the housing.
7. Remove any adhesive residue from the threads on the housing.
When doing so, note:
• No dirt must get into the housing during cleaning.
• A brass brush must be used for the external thread.
• A thread tap must be used for the internal thread.
15.15.2
Fitting the O
2
coupling
1. Pick up the O
2
coupling
.
2. Apply Loctite
to the thread of the O
2
coupling
.
3. Tightly screw the reducer into the O
2
coupling
by hand.
When doing so, note: Loctite
must harden according to
the manufacturer's specifications.
4. Apply Loctite
to the thread of the reducer.
Risk due to adhesive in the gas flow!
If the first two threads of the O
2
coupling are moistened with
adhesive, this adhesive can get into the gas flow to the patient
and lead to airway irritation.
Do
not
moisten the first two threads with adhesive.