Repairs to OXYWAY Fast II
15.30.2
Assembling the cover assembly
1. Place the large O-ring
in the groove in the multistep flow
selector
2. Place the small O-ring
in the groove in the multistep flow
selector
3. Use the Allen key to place the O-ring
in the groove in the
housing cover
4. Use the brush to apply a thin coat of oxygen-compatible
lubricant
to the O-ring
in the housing cover
.
5. Place the housing cover
in the cover clamping device
6. Tighten the cover clamping device
in the vise.
7. Place the multistep flow selector
on the housing cover
8. Pull the union nut
over the multistep flow selector
and
onto the housing cover
and screw on by hand.
9. Remove the adhesive residue from the threads on the multistep
flow selector
.
When doing so, note:
• No dirt must get into the housing during cleaning.
• A brass brush must be used for the external thread.
• A thread tap must be used for the internal thread.
10.Pick up the inhalation outlet