Repairs to OXYWAY Fine I
11.Fit the pressure gauge cap as protection against impact and
falling
(see "16.7.2 Fitting the pressure gauge cap as
protection against impact and falling", page 179)
.
12.Fit the O-ring
in the connecting bolt
(see "16.8.2 Fitting the O-ring in the connecting bolt", page 180)
16.26 Replacing the flow indicator
Required tools
• Counter tool with fastening handle
• Watchmaker's screwdriver
• Vise
16.26.1
Removing the flow indicator
Requirements
• The O-ring
has been removed from the connecting bolt
(see "16.8.1 Removing the O-ring from the connecting bolt",
page 179)
• The pressure gauge cap as protection against impact and
falling
has been removed
(see "16.7.1 Removing the pressure
gauge cap as protection against impact and falling", page 179)
1. Clamp the bottom part of the counter tool
2. Put the top part of the counter tool
onto the bottom part of the
counter tool
3. Screw the hand-screwed connection
of the pressure reducer
onto the bottom part of the counter tool
by hand.
4. Tightly screw the hand-screwed connection
of the counter tool
.
5. Use the open-end wrench to loosen the flow indicator
and
then remove it.
6. Use the watchmaker's screwdriver
to carefully remove the
two sealing disks
from the threaded hole of the flow
indicator
.
When doing so, note:
• The sealing surface in the housing must not be damaged.
• Ensure that no particles of dust get into the pressure reducer.