Repairs to OXYWAY Click
1. Lightly tension the cover clamping device
in the vise.
2. Insert the pressure reducer with O
2
sampling point
into the
cover clamping device
.
When doing so, note:
• The O
2
sampling point
must be facing downwards.
• The cover clamping device
must clamp onto the housing
cover
3. Tighten the cover clamping device
in the vise.
4. Use the open-end wrench to loosen the inhalation outlet
from the multistep flow selector
.
5. Remove the inhalation outlet
from the multistep flow
selector
6. Remove the adhesive residue from the threads on the multistep
flow selector
.
When doing so, note:
• No dirt must get into the housing during cleaning.
• A brass brush must be used for the external thread.
• A thread tap must be used for the internal thread.
17.7.2
Fitting the inhalation outlet
1. Pick up the inhalation outlet
.
2. Apply Loctite
to the thread of the inhalation outlet
.
3. Insert the inhalation outlet
into the multistep flow selector
4. Use the torque wrench to tightly screw in the inhalation
outlet
in the multistep flow selector
.
When doing so, note: Loctite
must harden according to
the manufacturer's specifications.
5. Loosen the vise.
Risk due to adhesive in the gas flow!
If the first two threads of the inhalation outlet are moistened with
adhesive, this adhesive can get into the gas flow to the patient and
lead to airway irritation.
Do
not
moisten the first two threads with adhesive.