Repairs to OXYWAY Click
17.11.2
Fitting the threaded pin in the multistep flow
selector
1. Insert the threaded pin
into the multistep flow selector
and
screw in using the Allen key.
When doing so, note: The threaded pin
must not be screwed
into the hole for the orifice plate assembly
2. Assemble the multistep flow selector assembly
(see "17.9.2 Assembling the multistep flow selector assembly",
page 213)
.
3. Fit the multistep flow selector assembly
(see "17.8.2 Fitting the multistep flow selector assembly",
page 211)
.
17.12 Replacing the multistep flow selector
17.12.1
Removing the multistep flow selector
Requirement
• The multistep flow selector assembly
has been removed
(see "17.8.1 Removing the multistep flow selector assembly",
page 211)
.
• The multistep flow selector assembly
(see "17.9.1 Disassembling the multistep flow selector assembly",
page 212)
.
• The threaded pin
has been removed from the multistep flow
selector
(see "17.11.1 Removing the threaded pin from the
multistep flow selector", page 216)
1. Dispose of the multistep flow selector
.
17.12.2
Fitting the multistep flow selector
1. Pick up the multistep flow selector
2. Fit the threaded pin
in the multistep flow selector
(see "17.11.2 Fitting the threaded pin in the multistep flow
selector", page 217)
3. Assemble the multistep flow selector assembly
(see "17.9.2 Assembling the multistep flow selector assembly",
page 213)
.
4. Fit the multistep flow selector assembly
(see "17.8.2 Fitting the multistep flow selector assembly",
page 211)
.