EN
233
Repairs to OXYWAY Click
10.Clamp the cover clamping device
onto the housing
11.Tighten the cover clamping device
in the vise.
12.Use the open-end wrench to loosen the O
2
sampling point
from the housing
and then remove it.
13.Fit the O
2
sampling point
.
17.28 Replacing the disposable filter
17.28.1
Removing the disposable filter
Requirement
• The O
2
sampling point
has been removed
• The housing
has been removed
(see "17.27.1 Removing the housing", page 231)
1. Dispose of the disposable filter
.
17.28.2
Fitting the disposable filter
1. Pick up the disposable filter
2. Fit the housing
(see "17.27.2 Fitting the housing", page 232)
.
3. Fit the O
2
sampling point
.