Test report
According to the service manual WM 66611
WEINMANN Emergency Medical Technology GmbH + Co. KG, Frohbösestraße 12, 22525 Hamburg, GERMANY
Tel: +49 40 88 18 96-120, Fax: +49 40 88 18 96-481
WM 66611 – 07/
2021
Page 2 of 3
Serial number:
8.2 Test equipment required
OK
Not
OK
Test equipment
(according to service manual OXYWAY pressure reducer WM 66611)
8.3
Preparing the final inspection
OK
Not
OK
x
The pressure reducer shows no signs of external damage.
x
The labels on the pressure reducer are present and legible.
x
The embossed serial number is legible.
x
The connection threads show no sign of external damage and turn smoothly.
8.4
Checking the system for leaks
OK
Not
OK
x
No bubbles form between the connecting bolt and the housing.
x
No bubbles form between the contents indicator and the housing.
x
No bubbles form on the safety valve.
x
No bubbles form at the small opening between the housing and the housing cover.
x
No bubbles form between the multistep flow selector and the inhalation outlet.
8.5
Checking the contents indicator
OK
Not
OK
x
The needle on the contents indicator is at "0".
x
The contents indicator can be rotated.
x
The contents indicator displays the cylinder pressure.
x
The needle of the contents indicator moves towards "0" without catching.
8.6
Checking the static outlet pressure p4
OK
Not
OK
Cylinder pressure p1 (from the table in Chapter 8.6)
bar
x
The static outlet pressure p4 is within the tolerance range
(from the table in Chapter 8.6).
bar
Static outlet pressure p4 is constant after 1 min.
8.7
Checking the constant output (flow) at the inhalation outlet
OK
Not
OK
Cylinder pressure p1 (100 bar)
bar
OXYWAY Fast III
Pressure reducer with maximum output 6 l/min
x
0.5 l/min ± 0.11 l/min
l/min
x
1 l/min ± 0.26 l/min
l/min
x
1.5 l/min ± 0.41 l/min
l/min
x
2 l/min ± 0.36 l/min
l/min
x
2.5 l/min ± 0.45625 l/min
l/min
x
3 l/min ± 0.5475 l/min
l/min
x
4 l/min ± 0.73 l/min
l/min
x
5 l/min ± 0.9125 l/min
l/min
x
6 l/min ± 1.095 l/min
l/min
x
Turn from 6 l/min to 3 l/min: 3 l/min ± 0.5475 l/min
l/min
x
Turn from 3 l/min to 0 l/min: 0 l/min ± 0.04 l/min
l/min
Summary of Contents for OXYWAY Click