Final inspection OXYWAY Fine I
10.6 Checking the flow indicator
1. Screw the hand-screwed connection
onto the valve of
the oxygen cylinder.
2. Screw the hand-screwed connection
tight by hand.
3. Close the multistep flow selector
4. Check the needle on the flow indicator
.
Requirement:
The needle on the flow indicator
is at "0".
5. Open the valve of the oxygen cylinder slowly.
6. Fully open the multistep flow selector
7. Close the valve on the oxygen cylinder.
Requirement:
The needle of the flow indicator
moves towards "0"
without catching.
8. Close the multistep flow selector
9. Remove the hand-screwed connection
from the valve of
the oxygen cylinder.
10.7 Checking the static outlet
pressure p4
1. Screw the hand-screwed connection
onto the valve of
the oxygen cylinder.
2. Screw the hand-screwed connection
tight by hand.
3. Close the multistep flow selector
4. Screw the adapter
onto the inhalation outlet
5. Screw the union nut of the supply test set
onto the
adapter