80
EN
OXYWAY
WM 66
611 07/
202
1
Repairs to OXYWAY Fix I, OXYWAY Fix III
1. Remove the contents indicator
from the housing.
2. Remove the O-ring
from the contents indicator
13.16.2
Fitting the contents indicator
1. Use the brush to wet the groove in the contents indicator
with oxygen-compatible lubricant
.
2. Position the O-ring
on the contents indicator
3. Insert the contents indicator
into the housing.
4. Fit the threaded locking pin
(see "13.15.2 Fitting the threaded locking pin", page 79)
5. Fit the spring for the piston
(see "13.13.2 Fitting the spring for the piston", page 77)
13.17 Replacing the O-ring in the contents
indicator
Required tools
Watchmaker's screwdriver
13.17.1
Removing the O-ring from the contents
indicator
Requirement
• The spring for the piston
(see "13.13.1 Removing the spring for the piston", page 77)
.
• The threaded locking pin
has been removed
(see "13.15.1 Removing the threaded locking pin", page 79)
• The contents indicator
has been removed
(see "13.16.1 Removing the contents indicator", page 79)
.
1. Use the watchmaker's screwdriver
to carefully remove
the O-ring
from the housing.