WM 66
611 07/
202
1
OXYWAY
EN
81
Repairs to OXYWAY Fix I, OXYWAY Fix III
13.17.2
Fitting the O-ring in the contents indicator
1. Position the O-ring
on the connecting screw on the contents
indicator
2. Fit the contents indicator
(see "13.16.2 Fitting the contents indicator", page 80)
.
3. Fit the threaded locking pin
(see "13.15.2 Fitting the threaded locking pin", page 79)
4. Fit the spring for the piston
(see "13.13.2 Fitting the spring for the piston", page 77)
.
13.18 Replacing the housing cover
13.18.1
Removing the housing cover
Requirement
• The spring for the piston
has been removed
(see "13.13.1 Removing the spring for the piston", page 77)
• The piston
has been removed
(see "13.14.1 Removing the piston", page 78)
1. Dispose of the housing cover
13.18.2
Fitting the housing cover
1. Pick up the housing cover
2. Insert the piston
into the housing cover
(see "13.14.2 Fitting the piston", page 78)
.
3. Fit the spring for the piston
(see "13.13.2 Fitting the spring for the piston", page 77)
.
4. Attach the label
(see "13.3.2 Attaching a label", page 66)
Back
Forward
Previous page
Next page