Repairs to OXYWAY Fast I, OXYWAY Fast I+, OXYWAY Fast III
14.4.2
Attaching a label
1. Pull the protective film off the back of the label
.
When doing so, note: Do not touch the adhesive surface of
the label
2. Stick the label
on the pressure reducer at the previously used
position.
When doing so, note: The adhesive surface must be free from dust
and grease.
14.5 Replacing the pressure screw in the
inhalation outlet
Required tools
14.5.1
Removing the pressure screw from the
inhalation outlet
1. Use the slot-head screwdriver to loosen the pressure screw
in the inhalation outlet
2. Remove the pressure screw
from the inhalation outlet
.
14.5.2
Fitting the pressure screw in the inhalation
outlet
1. Insert the pressure screw
into the inhalation outlet
When doing so, note: The slot on the pressure screw
must
be facing upwards when the screw is fitted.
2. Use the slot-head screwdriver to tightly screw in the pressure
screw
in the inhalation outlet
Back