94
EN
OXYWAY
WM 66
611 07/
202
1
Repairs to OXYWAY Fast I, OXYWAY Fast I+, OXYWAY Fast III
14.10.2
Fitting the hand-screwed connection
1. Pick up the hand-screwed connection
.
2. Fit the connecting bolt
(see "14.9.2 Fitting the connecting bolt", page 93)
3. Fit the O-ring
in the connecting bolt
(see "14.7.2 Fitting the O-ring in the connecting bolt", page 91)
.
14.11 Replacing the rib filter
Required tools
• Watchmaker's screwdriver
• Counter tool with fastening handle
• Counter cap
• Vise
14.11.1
Removing the rib filter
Requirement
The O-ring
has been removed from the connecting bolt
(see "14.7.1 Removing the O-ring from the connecting bolt",
page 90)
.
1. Clamp the bottom part of the counter tool
2. Put the top part of the counter tool
the counter tool
3. Screw the hand-screwed connection
of the pressure reducer
onto the bottom part of the counter tool
4. Tightly screw the hand-screwed connection
of the counter tool
.
5. Push the counter cap
over the safety valve and the housing.
6. Place the torque wrench onto the square of the counter cap
.
7. Use the torque wrench to loosen the housing from the connecting
bolt
.
8. Remove the torque wrench and counter cap
9. Unscrew the housing from the connecting bolt
by hand and
then remove.
10.Use the watchmaker's screwdriver
to remove the rib filter
from the connecting bolt
.
Back
Forward
Previous page
Next page
Summary of Contents for OXYWAY Click