Repairs to OXYWAY Fix I, OXYWAY Fix III
13.15 Replacing the threaded locking pin
Required tools
13.15.1
Removing the threaded locking pin
Requirement
The spring for the piston
(see "13.13.1 Removing the spring for the piston", page 77)
1. Use the Allen key to loosen the threaded locking pin
the housing.
2. Remove the threaded locking pin
from the housing.
13.15.2
Fitting the threaded locking pin
1. Insert the threaded locking pin
into the housing.
2. Use the Allen key to lightly screw in the threaded locking pin
3. Fit the spring for the piston
(see "13.13.2 Fitting the spring for the piston", page 77)
.
13.16 Replacing the contents indicator
Required tools
• Brush
13.16.1
Removing the contents indicator
Requirement
• The spring for the piston
has been removed
(see "13.13.1 Removing the spring for the piston", page 77)
• The threaded locking pin
(see "13.15.1 Removing the threaded locking pin", page 79)