Overview
Welch Allyn CP 100 and CP 200 Electrocardiograph
Battery Life Test
This displays the percentage of battery life remaining on the battery, based on the number
of charges it has incurred. This percentage approximates the life left in the battery based
on this information.
Printer Test
Caution
Only reset when the battery has been replaced and charged for at
least 12hrs continuously.
Function
Description
Figure 24. “Battery Life Test” Screen
Reset
Resets the battery life to 100%. This is used only
when the battery has been replaced
Reset Pages
Resets page count for the number of prints made on
the printer. Should only be reset when print head has
been replaced.
9:17AM Mar 16 05
Battery Life Test
Battery Life Remaining: 99%
Reset
Reset
Pages
Exit
Function
Description
Figure 25. Printer Test” Screen
Test Page
Prints a test page which repeats until the “Cancel”
button is pressed. Refer to Functional Verification test
for explanation. See
Ramp Test
Prints a one page ramp test. Refer to Functional
Verification test for explanation. See
.
9:17AM Mar 16 05
Printer Test
Select a print test
Test Page
Ramp
Page
Exit
Summary of Contents for CP 100
Page 6: ...vi Contents Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 7: ...1 1 Safety Summary Introduction 2 Symbols 2 Servicing the Electrocardiograph Safely 4 ...
Page 34: ...28 Overview Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 47: ...4 41 Troubleshooting Problem Solving Suggestions 42 Limited Warranty 45 Service Policy 46 ...
Page 82: ...76 Field Replaceable Parts Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 90: ...84 Appendix C Software Updates Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 91: ...D 85 Interconnect Diagrams CP 200 Interconnect Diagram 86 CP 100 Interconnect Diagram 87 ...
Page 93: ...Service Manual Appendix D Interconnect Diagrams 87 CP 100 Interconnect Diagram ...
Page 94: ...88 Appendix D Interconnect Diagrams Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 95: ......