26
Overview
Welch Allyn CP 100 and CP 200 Electrocardiograph
MMCSDTEST.EXE:
Used to test the SD memory card write/read function of the unit. Can also be used to test an SD memory card
for problems.
GO
: Starts the SD memory card test by pressing “G”
STOP:
Stops the SD memory card test by pressing “S”
ERASE ALL:
Erases all files written by this tester by pressing “E”
Card Present:
If checked the unit detects the SD memory card inserted.
Write Protect:
If checked the unit detects the lock is enabled on the SD memory card.
Pulse:
Flashes when test is active
Note:
Once the test is started it runs continuously. Press the “S” key to stop the test.
NORMANDY.EXE:
Used to return to the ECG program.
NORMTESTER.EXE
Strictly used for development purposes.
PPPTEST.EXE
Strictly used for development purposes.
PPPTEST2.EXE
Used to see battery and charger status/state. There are a number of selections on this screen. Listed below
are the most commonly used functions. To activate a selection press the letter that coincides with the
selection.
Real-time:
Activates a real-time monitoring of the readings from the battery circuit.
Battery Voltage:
Unplugged from AC this reflects the Voltage of the battery in millivolts; plugged into
AC this reflects the charge voltage in millivolts.
Fast charge = 6800 – 7300
Trickle charge = 6300 – 6800
Battery Current:
Indicates charge current in mA when charger is on.
Charger State:
Indicates if charge circuit is on or off
Charge State:
Indicates if the Fast Charge is on or off
Fast Charge On:
Turns fast charge on. If battery is fully charged the sensing circuit can work against
you. It only changes the reading for a brief second before switching back.
Fast Charge Off:
Turns fast charge off.
Charger On:
Turns charger on when plugged into AC.
Charger Off:
Turns charger off.
Function Description
Summary of Contents for CP 100
Page 6: ...vi Contents Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 7: ...1 1 Safety Summary Introduction 2 Symbols 2 Servicing the Electrocardiograph Safely 4 ...
Page 34: ...28 Overview Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 47: ...4 41 Troubleshooting Problem Solving Suggestions 42 Limited Warranty 45 Service Policy 46 ...
Page 82: ...76 Field Replaceable Parts Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 90: ...84 Appendix C Software Updates Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 91: ...D 85 Interconnect Diagrams CP 200 Interconnect Diagram 86 CP 100 Interconnect Diagram 87 ...
Page 93: ...Service Manual Appendix D Interconnect Diagrams 87 CP 100 Interconnect Diagram ...
Page 94: ...88 Appendix D Interconnect Diagrams Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 95: ......