Service Manual
Functional Verification
39
4.
Verify the electrocardiograph passed the test. (If the unit current never exceeds
5.00mA, the shielding is sufficient.)
Hi-Pot Test Connections
Chassis, Earth and Patient Leakage Current Tests
Verify continued electrical safety of the device, using IEC 60601-1 or ANSI/AAMI ES1
methods and limits. Test for the following:
•
Earth leakage
•
Chassis leakage
•
Patient leakage
Cable
Electrocardiograph Connection
Hi-Pot Tester Connection
Connector/
Cable
Connect to
Wiring
Connect to
Power Cord
IEC Power cord
AC Input
Connector
Hot and neutral (black and white) wires connected together and
terminated appropriately for your hi-pot tester. Cut and terminate
the ground (green) wire to prevent possible injury.
Return
Note
Refer to the Electrical Safety Analyzer manual for complete details on set-up and
use for testing.
Summary of Contents for CP 100
Page 6: ...vi Contents Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 7: ...1 1 Safety Summary Introduction 2 Symbols 2 Servicing the Electrocardiograph Safely 4 ...
Page 34: ...28 Overview Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 47: ...4 41 Troubleshooting Problem Solving Suggestions 42 Limited Warranty 45 Service Policy 46 ...
Page 82: ...76 Field Replaceable Parts Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 90: ...84 Appendix C Software Updates Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 91: ...D 85 Interconnect Diagrams CP 200 Interconnect Diagram 86 CP 100 Interconnect Diagram 87 ...
Page 93: ...Service Manual Appendix D Interconnect Diagrams 87 CP 100 Interconnect Diagram ...
Page 94: ...88 Appendix D Interconnect Diagrams Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 95: ......