44
Troubleshooting
Welch Allyn CP 100 and CP 200 Electrocardiograph
Table 5. System Messages (alphabetical order)
System Message
Problems
Actions
“A test is being printed that has not been
saved and will be lost. Continue
shutdown?” (CP 200)
was pressed while printing an
Auto ECG.
To shut down without saving the test, press
.
To cancel the shutdown, press
.
“Audit Trail is too large. Please print and
purge.” (CP 200)
Audit information is approaching its
maximum allotted storage.
Print and purge the audit trail. See “Working With the
Audit Trail”.
“Battery extremely low. Please connect
AC power.”
Battery low or bad.
•
Recharge the battery. See “Recharging a Fully
Discharged Battery” in the Directions for Use
manual.
•
Replace battery. See
“Insufficient space available”
Not enough space on memory card.
•
Delete some tests from the card at a PC.
•
Use a different card.
“Memory card error”
Problem writing to memory card.
•
Verify that the write-protect tab is in the
unprotected position.
•
Reseat the card in its slot.
•
Use a different card.
“Out Of Paper”
•
Printer is out of paper.
•
Printer door is open.
•
Load paper. See “Loading the Thermal Chart
Paper” in the Directions for Use.
•
Close the printer door. See “Loading the Thermal
Chart Paper” in the Directions for Use.
“Paper Error” (CP 200)
“Paper Jam” ( CP 100)
Paper was loaded incorrectly.
•
Reload the paper. See “Loading the Thermal
Chart Paper” in the Directions for Use.
“Powering down”
Low battery.
Recharge the battery. See “Recharging a Fully
Discharged Battery”.
“Problem loading the following settings:
<System>
<ECG>
*
*
*
Using default settings.”
Problem loading your settings at startup.
May indicate memory problems.
Contact Technical Support. For phone numbers, see
.
“Shutdown?”
was pressed while printing a
Rhythm ECG.
To shut down, press
.
To cancel the shutdown, press
.
“Temperature Error”
Printer head temp is too high.
Allow to cool, then try again.
“Test has not been saved. Continue
shutdown?” (CP 200)
was pressed after printing an
Auto ECG.
To shut down without saving the test, press
.
To cancel the shutdown, press
.
“Unable to communicate with
workstation“ (CP 200)
•
Configuration not correct.
•
Cable not properly connected.
•
Verify communication settings on the PC.
•
Check cable connections.
Summary of Contents for CP 100
Page 6: ...vi Contents Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 7: ...1 1 Safety Summary Introduction 2 Symbols 2 Servicing the Electrocardiograph Safely 4 ...
Page 34: ...28 Overview Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 47: ...4 41 Troubleshooting Problem Solving Suggestions 42 Limited Warranty 45 Service Policy 46 ...
Page 82: ...76 Field Replaceable Parts Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 90: ...84 Appendix C Software Updates Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 91: ...D 85 Interconnect Diagrams CP 200 Interconnect Diagram 86 CP 100 Interconnect Diagram 87 ...
Page 93: ...Service Manual Appendix D Interconnect Diagrams 87 CP 100 Interconnect Diagram ...
Page 94: ...88 Appendix D Interconnect Diagrams Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 95: ......