Service Manual
CP 100 and CP 200 Electrocardiograph
59
5.
Unplug the two battery connectors. See
•
You may need to pull hard. Use pliers
gently if necessary.
6.
Connect a new battery by matching
connector sizes.
Do not force connectors to mismatch.
7.
Place the battery into the battery
compartment. Replace the battery door
and screws. Turn the electrocardiograph
right-side-up.
It turns on automatically and displays a
prompt to enter date and time.
8.
Reset the unit after you replace the battery.
See
Battery Accuracy Test (page 31)
for
directions.
9.
Enter the date and time.
The electrocardiograph is ready to use.
10. Discard the old battery appropriately.
Figure 49. Battery Connectors
Procedure
Battery
connectors
The battery is a sealed, lead-acid battery and must be recycled or
disposed of properly according to national or local regulations.
Summary of Contents for CP 100
Page 6: ...vi Contents Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 7: ...1 1 Safety Summary Introduction 2 Symbols 2 Servicing the Electrocardiograph Safely 4 ...
Page 34: ...28 Overview Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 47: ...4 41 Troubleshooting Problem Solving Suggestions 42 Limited Warranty 45 Service Policy 46 ...
Page 82: ...76 Field Replaceable Parts Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 90: ...84 Appendix C Software Updates Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 91: ...D 85 Interconnect Diagrams CP 200 Interconnect Diagram 86 CP 100 Interconnect Diagram 87 ...
Page 93: ...Service Manual Appendix D Interconnect Diagrams 87 CP 100 Interconnect Diagram ...
Page 94: ...88 Appendix D Interconnect Diagrams Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 95: ......