66
Disassembly and Repair Procedures
Welch Allyn CP 100 and CP 200 Electrocardiograph
Replacing the Keyboard Assembly (CP 200)
Procedure
1.
Open unit using the
directions.
2.
Disconnect the ground and the flat data
cable to the LCD and the flat data cable,
the cables to the Inverter, to the Keypad
Assembly.
3.
Remove the 13 screws holding in the
Keyboard Assembly.
4.
Gently lift out the Keyboard Assembly.
Reassembly Note
1.
Install and tighten the 3 screws first
(shown in
) to align the keyboard.
DO NOT tighten the screws past the
specified torque.
2.
Install remaining 10 screws using the
specified torque.
Figure 56. Replacing the Keyboard Assembly (CP 200)
Cable between Inverter
board and Keyboard
Flat Cable from the
LCD to Keyboard
Screws
LCD Assembly
Keypad
Assembly
Summary of Contents for CP 100
Page 6: ...vi Contents Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 7: ...1 1 Safety Summary Introduction 2 Symbols 2 Servicing the Electrocardiograph Safely 4 ...
Page 34: ...28 Overview Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 47: ...4 41 Troubleshooting Problem Solving Suggestions 42 Limited Warranty 45 Service Policy 46 ...
Page 82: ...76 Field Replaceable Parts Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 90: ...84 Appendix C Software Updates Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 91: ...D 85 Interconnect Diagrams CP 200 Interconnect Diagram 86 CP 100 Interconnect Diagram 87 ...
Page 93: ...Service Manual Appendix D Interconnect Diagrams 87 CP 100 Interconnect Diagram ...
Page 94: ...88 Appendix D Interconnect Diagrams Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 95: ......