32
Functional Verification
Welch Allyn CP 100 and CP 200 Electrocardiograph
4.
Turn fast charge ON. Check DVM for 6.8 to 7.3 VDC.
5.
Turn fast charge off. Check DVM for 6.3 to 6.8 VDC.
6.
Unplug the unit from the AC power.
7.
Disconnect the DVM and install the battery.
LED Test
Verify the Power LED light is on when unit is plugged in, and verify the light is off when
unit is unplugged.
1.
Plug the unit into AC power.
2.
Verify the AC indicator LED under the On/Standby button is lit.
3.
Unplug the AC power cord.
4.
Verify the LED is off.
5.
Plug the unit back into AC power.
LCD Test
CP 200
1.
Choose
Main Menu
>
Systems Setting
>
Device info
>
Service Info
>
LCD Test.
2.
Verify the test pattern is visible without breaks and is consistent over the entire
screen.
3.
Verify Contrast can be adjusted to adequately see the display.
CP 100
1.
Disconnect the patient cable.
2.
Turn
ON
unit. The patient hook up screen appears.
3.
Verify the display is visible without breaks.
Set Unit Initial Configuration/SD Slot Test
Initial Configuration of Unit
1.
Verify the Serial Number on the unit is the same as displayed on the unit’s screen. If
incorrect, enter correct S/N. See
on page 16, Service
Info.
Note
There may be a few seconds delay in the LED response to the line being removed
and added.
Note
If contrast adjustment is needed refer to the Directions for Use manual for
CP 100.
Summary of Contents for CP 100
Page 6: ...vi Contents Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 7: ...1 1 Safety Summary Introduction 2 Symbols 2 Servicing the Electrocardiograph Safely 4 ...
Page 34: ...28 Overview Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 47: ...4 41 Troubleshooting Problem Solving Suggestions 42 Limited Warranty 45 Service Policy 46 ...
Page 82: ...76 Field Replaceable Parts Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 90: ...84 Appendix C Software Updates Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 91: ...D 85 Interconnect Diagrams CP 200 Interconnect Diagram 86 CP 100 Interconnect Diagram 87 ...
Page 93: ...Service Manual Appendix D Interconnect Diagrams 87 CP 100 Interconnect Diagram ...
Page 94: ...88 Appendix D Interconnect Diagrams Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 95: ......