38
Functional Verification
Welch Allyn CP 100 and CP 200 Electrocardiograph
Dielectric Strength Test
It is important to complete the dielectric strength test after any procedure in which the
monitor is opened to verify proper patient isolation.
The patient isolation test requires an AC Withstand Voltage (hi-pot) Tester, such as the AR
3605 or equivalent. If this equipment is not available, Welch Allyn can perform the patient
isolation test for you quickly, for a nominal fee
.
To create test cables for the patient isolation test. See
Set up the test parameters on the AC Withstand Voltage Tester as follows:
To create test cables for this procedure. See
1.
Connect the modified power cord between the return jack on the tester and the AC
input of the unit.
2.
Connect the high voltage cable between the equipotential stud and the high voltage
on the tester.
3.
Press the test button to run the AC withstand voltage test.
WARNING
Failure to run the dielectric test when indicated could cause serious
injury to patients, and could lead to damage to the monitor.
WARNING
Always follow the tester manufacturer’s operation instructions
exactly. Failure to perform this test properly can result in serious injury or death.
WARNING
This test must be performed by qualified service personnel only.
WARNING
Run this test, only, on an insulated table top, and away from other
people and equipment.
WARNING
Must avoid connector shell and pe ground because pe is only
resistant to 500 Vac.
WARNING
The patient isolation test involves exposure to extremely high
voltages and is extremely hazardous.
Note
Refer to the tester manufacturer’s operation manual for complete details on set
up and use.
Parameter
Specification
Voltage
1500Vac
Maximum Limit
5.00mA
Minimum Limit
0.15 mA
Ramp Up
10 s
Dwell
1 min.
Delay
0.5 s
Ramp Down
0.0 s
Arc Sense
0
Frequency
Local line (50 Hz or 60 Hz)
Continuity
Off
Connect
Off
Summary of Contents for CP 100
Page 6: ...vi Contents Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 7: ...1 1 Safety Summary Introduction 2 Symbols 2 Servicing the Electrocardiograph Safely 4 ...
Page 34: ...28 Overview Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 47: ...4 41 Troubleshooting Problem Solving Suggestions 42 Limited Warranty 45 Service Policy 46 ...
Page 82: ...76 Field Replaceable Parts Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 90: ...84 Appendix C Software Updates Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 91: ...D 85 Interconnect Diagrams CP 200 Interconnect Diagram 86 CP 100 Interconnect Diagram 87 ...
Page 93: ...Service Manual Appendix D Interconnect Diagrams 87 CP 100 Interconnect Diagram ...
Page 94: ...88 Appendix D Interconnect Diagrams Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 95: ......