50
Disassembly and Repair Procedures
Welch Allyn CP 100 and CP 200 Electrocardiograph
Replacing the Printer Assembly
Procedure
1.
Open unit using the
directions.
2.
Open the paper door and remove
all
paper
in the paper tray.
3.
Insert a flat blade screwdriver into the slit
opening located on the side of the auxiliary
paper cover.
4.
Apply outward pressure on the screwdriver
while pressing upward on the side of the
cover. Repeat process for the other side of
the cover. Lift the cover out of unit.
Figure 33. Inserting a Flat blade Screwdriver
5.
Remove the 4 screws in the bottom of the
printer tray. See
.
Figure 34. Removing the Screws in the Paper Tray
Auxiliary Paper Cover
Slit Openings
QuickStep
OperatingInstructions
StatECG
StatECG
Holddown
for1
second
AutomaticECG
Select
EnterPatient
Data
Select
DONE
Verify
ECGquality
Select
RhythmECG
Select
toend
Select
Changesettings
asdesired
ECG
IEC
C1
C2
C3
C4
C5
C6
R
L
N
F
AAM
I
V1
V2
V3
V4
V5
V6
RA
LA
RL
LL
VI
V2
V3
R4
V4
V5
V6
LA
LL
RL
4 Screws
Summary of Contents for CP 100
Page 6: ...vi Contents Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 7: ...1 1 Safety Summary Introduction 2 Symbols 2 Servicing the Electrocardiograph Safely 4 ...
Page 34: ...28 Overview Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 47: ...4 41 Troubleshooting Problem Solving Suggestions 42 Limited Warranty 45 Service Policy 46 ...
Page 82: ...76 Field Replaceable Parts Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 90: ...84 Appendix C Software Updates Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 91: ...D 85 Interconnect Diagrams CP 200 Interconnect Diagram 86 CP 100 Interconnect Diagram 87 ...
Page 93: ...Service Manual Appendix D Interconnect Diagrams 87 CP 100 Interconnect Diagram ...
Page 94: ...88 Appendix D Interconnect Diagrams Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 95: ......