Service Manual
CP 100 and CP 200 Electrocardiograph
53
6.
Turn the paper tray assembly upright.
7.
Slide the paper tray cover in half way until
the paper guide drops into the paper tray.
8.
Slide the paper tray cover out to remove it.
Figure 39. Removing the Paper Tray Cover
Reassembly Note
1.
Slide the paper tray cover half way into the
paper tray assembly.
2.
Place the paper guide under the roller on the
paper tray cover.
3.
While holding the paper guide in place, slide
the paper tray cover out until it hits the stop.
4.
Turn the paper assembly up-side-down.
5.
Replace the 3 screws in the holes located in
the bottom of the paper tray assembly. See
.
Procedure
Quick
StepOperating
Instructions
StatECG
StatECG
Holddown
for1
second
AutomaticECG
Select
EnterPatient
Data
Select
DO
NE
Verify
ECG
quality
Select
RhythmE
CG
Select
toend
Select
Changesettings
asdesired
EC
G
IEC
C1
C2
C3
C4
C5
C6
R
L
N
F
AAM
I
V1
V2
V3
V4
V5
V6
RA
LA
RL
LL
VI
V2
V3
R4
V4
V5
V6
LA
LL
RL
Paper Guide
Printhead
Direction to Slide Tray Cover to
Drop Paper Guide into the Tray.
Direction to Slide Paper Tray Cover to
Summary of Contents for CP 100
Page 6: ...vi Contents Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 7: ...1 1 Safety Summary Introduction 2 Symbols 2 Servicing the Electrocardiograph Safely 4 ...
Page 34: ...28 Overview Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 47: ...4 41 Troubleshooting Problem Solving Suggestions 42 Limited Warranty 45 Service Policy 46 ...
Page 82: ...76 Field Replaceable Parts Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 90: ...84 Appendix C Software Updates Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 91: ...D 85 Interconnect Diagrams CP 200 Interconnect Diagram 86 CP 100 Interconnect Diagram 87 ...
Page 93: ...Service Manual Appendix D Interconnect Diagrams 87 CP 100 Interconnect Diagram ...
Page 94: ...88 Appendix D Interconnect Diagrams Welch Allyn CP 100 and CP 200 Electrocardiograph ...
Page 95: ......