Service Manual
CP 100 and CP 200 Electrocardiograph
61
Replacing the Main Board
Procedure
1.
Open unit using the
directions.
2.
Remove printer assembly using the
Replacing the Printer Assembly (page 50)
directions.
3.
Disconnect the locking connection for the
power supply and the battery.
4.
Remove the ribbon cable that attaches the
top case.
5.
Remove the 5 self taping screws holding
the board into place. See
6.
Carefully lift and slide the board toward the
power supply.
Figure 51. Removing the Main Board
Ribbon Cable
Locking Connection
Power
Supply
Connection
Screws