Service Manual
CP 100 and CP 200 Electrocardiograph
65
Replacing the Function Keypad Assembly (CP 200)
Procedure
1.
Open unit using the
directions.
2.
Disconnect the cables to the LCD and the
Function Keyboard. See
.
3.
Remove the 5 screws holding the Function
Keypad Assembly. See
.
4.
Gently lift out the Function Keypad
Assembly.
Figure 55. Replacing the Keypad Assembly
LCD Assembly
Phillips Screws
Cable connection LCD
the and Function
Keypad