background image

10

3.203  Forced Oil Flushing Procedure
 

1. Check the oil level, if the oil level is well above the fi ll mark (This may indicate either the pump has been 
overfi lled with oil or has ingested a liquid or a large amount of vapor water or organic solvents). Please go to step 
2. If the oil level is even with the fi ll mark and you do NOT suspect corrosive gases or particulates (hence forth 
called contaminants) ingested have damaged the mechanism, run the pump for 15 minutes to allow the pump oil 
to warm up before going to step 2.

 

2. Turn off motor for the vacuum pump. Drain the oil into a clear plastic container (may need pliers to open drain 
valve).  Look for contaminations settling to the bottom of container.  If you see contaminants, you will need to 
repeat step 3 through 5 several times until the oil comes out clear.  The oil you drained from the pump came from 
the oil case only.  There may be contaminants in the pumping mechanism.  To be sure all contaminants have 
been removed, the pump mechanism needs to be fl ushed.

 

3. Make sure the belt guard is installed before proceeding further. Attach a short hose to the drain valve which 
runs into a clear plastic container.

 

4. Flushing the pump is carried out by adding a cup of DUOSEAL vacuum pump oil through the intake port (IN) 
while the pump is turned on for 15 to 20 seconds.  While adding the pump oil, the palm of your hand is placed 
lightly over the exhaust port (OUT). Look for water coming out of the drain. Turn off the pump.

 

5. Repeat step 4 until clean oil comes out of the drain hose.

 

6. Close the drain valve and fi ll the pump with the amount of DUOSEAL vacuum pump oil your pump needs.

 

7. Plug the intake (IN) port with a rubber stopper.  Turn the pump on and run the pump for 10 minutes.  Close the 
gas ballast.

 

8. Check the vacuum reading of the pump by connecting a thermocouple gauge tube to the pump’s intake.  If the 
pump is running nearly as good as when it was new, the total pressure reading you will read on wither of these 
two gauges will be at least 10 micron.

 

A simple way to connect the gauge tube to the pump is to run the threaded tip of tube through a hole in a rubber 
stopper.  Use pump oil as a lubricant for inserting the tube.  The stopper chosen should be bigger than the outer 
diameter of the intake fl ange.

3.204 Refi lling The Pump
 

After you are satisfi ed that the pump has been thoroughly fl ushed, refi ll the pump by pouring new DUOSEAL oil 
into the exhaust port.  Fill to the indicated level and start the pump with the intake closed.  A gurgling noise is 
characteristic when high pressure air is drawn through the pump.  It should disappear quickly as the pressure 
within the pump is reduced.  If gurgling continues, add suffi cient additional oil through the exhaust port until 
gurgling ceases.

3.30 

Shaft Seal Replacement

 

To replace the shaft seal of a pump, drain the oil and remove the pump pulley and key.  Remove the screws 
securing the old seal and pry it loose with a screwdriver or similar wedge, being careful not to mar the surface of 
the pump body against which the seal fi ts.  Discard the seal and its gasket, inspect all surfaces and repair any 
damages with a fi ne abrasive stone.  Wipe all sealing areas clean and place a fi lm of DUOSEAL oil on both the 
shaft and the inside bore of the new shaft seal.  Using a new gasket, carefully slide the new seal into position 
and center it on the shaft.  It is not necessary to apply any sealant to the gasket.  Tighten the mounting screws 
uniformly and refi ll the pump with DUOSEAL oil.  Follow instructions included in repair kit.

Summary of Contents for DUOSEAL 1405B-01

Page 1: ...essure will build up in the pump with the potential of the pump body bursting and causing possible injury to personnel in the area Part No 67 2061 R1 0 Printed in USA Welch Ilmvac 5621 W Howard Street...

Page 2: ...tions are found where failure to observe the instruction could result in damage to the equipment associated equipment and process These units conform to the SI International system of units of measure...

Page 3: ...10 3 30 Shaft Seal Replacement 10 3 40 Repairing Oil Leaks 11 3 50 Repairing Vacuum Leaks 11 3 60 Drive Problems 11 Section 04 Trouble Shooting Guide 4 10 Poor Vacuum Readings 12 4 20 Noisy Pump 12 4...

Page 4: ...to drive your particular pump at the recommended speed 1 13 Pump Location The pump should be located preferably in a clean and well ventilated area and adequate space should be provided wherever poss...

Page 5: ...creates an impedance to the flow of gas This is particularly true at low pressures in the millitorr range where the gas flow is substantially molecular in character The gas flow is then dependent upo...

Page 6: ...with the function and action of the pump which you have now required Remove the intake and exhaust port plugs and temporarily provide a stopper for the intake and a dust cap for the exhaust Review the...

Page 7: ...ir though the gas ballast valve to the mixture of vapor and gas being compressed the pressure required for delivery past the exhaust valve is reached with a considerably smaller reduction of the volum...

Page 8: ...rich in hydrogen and observing a sudden change in pressure on an electrical gauge The difference in calibration of these gauges for air and other gases will produce a distinct change in the pressure...

Page 9: ...system The required frequency of changes will vary with the particular system Experiences with the process will help you determine the normal period of operation before an oil change is required 3 103...

Page 10: ...rubber stopper Turn the pump on and run the pump for 10 minutes Close the gas ballast 8 Check the vacuum reading of the pump by connecting a thermocouple gauge tube to the pump s intake If the pump is...

Page 11: ...ng surfaces cleaned of any nicks 3 50 Repairing Vacuum Leaks 3 501 Surface Preparation Good Vacuum seals are an essential and important attribute of a good mechanical pump A good seal is dependent upo...

Page 12: ...t a forced oil flush 5 If noise level is still unsatisfactory consult with WELCH Technical Service at 847 676 8800 Ext 1 4 30 Excessive Oil Mist 1 Check oil level Maintain the level between the two li...

Page 13: ...cal Requirements 2 60Hz Electrical Requirements 3 1 Phase Replacement Parts Shaft Seal Lip Seal 1401E 01 Shaft Seal Mechanical Seal 1401E Air Filter Screen 41 0891 Intake Chamber Cover Gasket 41 1308...

Page 14: ...Section 6 DIMENSIONAL DRAWING 14...

Page 15: ...nge High RPM direct drive vacuum pumps on the other hand experience a rapid fall off of pumping speed below 100 microns A comparison of the DuoSeal pumping speed curves shown below versus high RPM pum...

Page 16: ...Section 8 EXPLODED VIEW 16...

Page 17: ...CER 12 3 2 61 0000 WASHER STEEL 3 16 42 1 41 2160 BAFFLE ASSEMBLY 13 10 2 63 0193 WASHER LOCK 1 4 MEDIUM 43 1 41 2175 SETSCREW 1 4 20 x 2 25 LG 14 1 4 06 0754 1 1 ROTOR CLIP 44 6 41 2363 WASHER 5 16 I...

Page 18: ...101 Parts List for Model 61 8101 Item Qty P N Description 1 2 61 2229A Oil Window Glass 2 1 61 2278 Drain Valve 3 2 61 2370 Oil Window Cover 4 1 61 8833 Oil Case Machined 5 4 62 1014 Screw 8 32x3 8 Fl...

Page 19: ...ust Filter Use with pump model Dimension Catalog Number Diameter Height Filter Replacement Element 1392 1399 1400 1376 1380 1402 1400 1397 1374 2 5 5 5 2 5 6 25 9 1417L 1417G 1417H 01 1417L 1417G 1417...

Page 20: ...1922 Cat No 1 Element must be ordered with trap A2 Distance between ISO NW inlet and outlet connections Dry Ice Cold Traps Removable 3 quarter center well for dry ice alcohol slurry do not use acetone...

Page 21: ...best of our knowledge the above listed component is not hazardous according to OSHA 1910 1200 or one or more state right to know lists SECTION 3 HAZARDS IDENTIFICATION Major Route of Entry Skin conta...

Page 22: ...ection in Section 8 of this MSDS Inhalation Vaporization is not expected at ambient temperatures This material is not expected toques inhalation related disorders underanticipated conditions of use In...

Page 23: ...in Section 8 and Disposal Considerations in Section 13 of this MSDS Do not touch damaged containers or spilled material unless wearing appropriate protective equipment Slipping hazard do not walk thr...

Page 24: ...splashing or spraying conditions are present Protective clothing may include long sleeve outer garment apron or lab coat If significant contact occurs remove oil contaminated clothing as soon as possi...

Page 25: ...e low acute and sub acute toxicities in animals Effects from single and short term repeated exposures to high concentrations of mineral oil mists well above applicable workplace exposure levels includ...

Page 26: ...ry This product and or its components are listed on the Toxic Substance Control Act TSCA Inventory SARA 302 304 The Superfund Amendments and Reauthorization Act of 1986 SARA Title III requires Facilit...

Page 27: ...arks SECTION 16 OTHER INFORMATION Revision 4 Revision Date 11 11 03 ABBREVIATIONS AP Approximately EQ Equal GT Grater Than LT Less Than NA Not Applicable ND No Data NE Not Established ACGIH American C...

Page 28: ...or possession does not convey any rights to loan sell or otherwise disclose said information reproduction or use of said information for any purpose other than that in connection with said information...

Reviews: