background image

9

Section 3:  MAINTENANCE

3.10 

Vacuum Problems

3.101  Pressure Determinations

 

Leakage, contamination and unusual outgassing are the general causes of problems associated with poor vacuum. 

  

To operate at maximum efficiency a system must be thoroughly clean.  If the system is completely clean and free 

 

from leaks, and unwarranted vacuum problems still exist, the pump should be checked.  A simple criterion for the 

 

condition of a mechanical pump is a determination of its ultimate pressure capability.  This can be accomplished 

 

by attaching a gauge directly to the pump.  The gauge may be any suitable type provided consideration if given 

 

to the limitations of the gauge being used.  Refer to Paragraph 1-8 for further suggestions.  If the pressure is 

 

unusually high, the pump may be badly contaminated, low on oil or malfunctioning.  On the other hand, if the 

 

pressure is only slightly higher than the guaranteed pressure of the pump, an oil change may be all that is required.

3.102  Oil Contamination

 

The most common cause of a loss in efficiency in a mechanical pump is contamination of oil.  It is caused by 

 

condensation of vapors and by foreign particles.  The undesirable condensate emulsifies with the oil which is 

 

recirculated and subject to re-evaporation during the normal cycle of pump activity thus reducing the ultimate 

 

vacuum attainable.  Some foreign particles and vapors may form a sludge with the oil, impair sealing and 

 

lubrication and cause eventual seizure.  A gas ballast calve is helpful in removing vapors, especially water, but it 

 

is not equally effective on all foreign substances; therefore, periodic oil changes are necessary to maintain 

 

efficient operation of the system.  The required frequency of changes will vary with the particular system,.  

 

Experiences with the process will help you determine the normal period of operation before an 

 

oil change is required.

3.103  Oil Overheating

 

This pump is designed to operate continuously below 10 Torr.  Continuous operation of this pump above 10 Torr 

 

will lead to overheating and eventual pump failure.  See Section 2.21, High Pressure Operation.

WARNING

   Solvents, such as acetone, used to clean different  pumps 

are extremely hazardous.  Inhalation may be fatal 

or harmful.  Acetone is also extremely flammable. 

Never use acetone near heat or open flame.  

Use only in a well ventilated area.

3.20 

Oil Changes and Oil Level

3.201  Developing a Maintenance Schedule

 

After studying many examples of pump failure, Welch has found the most common reason is poor condition of 

 

the oil.  This is why a regular maintenance schedule for the oil is critical to obtain the longest service life out of 

 

your DUOSEAL pump.  Welch recommends that you examine the condition of the oil on a daily basis in the early

  

days of a new process or experiment.  You want to look for discoloration of the oil and whether the oil level is 

 

rising.  The discoloration can indicate deterioration of the oil and a rising oil level can indicate condensation of 

 

vapors is occurring in the pump.  When changes occur, the oil needs to be change.  If no changes in the oil level 

 

or color are observed, extended pump service life is obtained if the oil is changed every three to four months.

3.202  Forced Oil Flushing

 

When you drain oil through the drain valve, you are not removing the oil and contaminants that are inside the 

 

pumping mechanism.  You are removing oil only from the oil case.  Welch recommends a forced oil flush of 

 

DUOSEAL pumps be performed at the regular maintenance oil change.  

 

The procedure for the forced oil flush is given below.

Summary of Contents for DUOSEAL 1399B-01

Page 1: ...sting and causing possible injury to personnel in the area Gardner Denver Thomas Inc Welch Vacuum Technology 5621 W Howard Street Niles IL 60714 Phone 847 676 8800 Fax 847 677 8606 Technical Support 847 588 2366 E Mail gdwelchvacuum gardnerdenver com Part No 67 2063 Web Page www welchvacuum com Printed in USA SAVE THESE INSTRUCTIONS Manufactured in an ISO 9000 Registered Environment 505 872 0037 i...

Page 2: ...ions are found where failure to observe the instruction could result in damage to the equipment associated equipment and process These units confirm to the SI International system of units of measurement The following symbols with recommendation of IEC1010 of warning will be found on the pump Caution refer to accompanying documents Caution risk of electrical shock Caution hot surface WARNING Motor...

Page 3: ...evel 10 3 30 Shaft Seal Replacement 10 3 40 Repairing Oil Leaks 11 3 50 Repairing Vacuum Leaks 11 3 60 Drive Problems 11 Section 04 Trouble Shooting Guide 4 10 Poor Vacuum Readings 12 4 20 Noisy Pump 12 4 30 Excessive Oil Mist 12 Section 05 Quick Reference Chart Chart 1 13 Chart 2 13 Section 06 Dimenional Drawings with rotation Vacuum Pump 14 Section 07 Pumping Speed Curves Speed Curves 15 Section...

Page 4: ...y to drive your particular pump at the recommended speed 1 13 Pump Location The pump should be located preferably in a clean and well ventilated area and adequate space should be provided wherever possible for routine maintenance such as changes of oil and belt adjustments and replacements Above all the pump should be located as closely as possible to its system in order to utilize it most efficie...

Page 5: ...he gas flow is substantially molecular in character The gas flow is then dependent upon the kinetic activity of the molecules to bring it to the intake of the pump 1 162 The Effects of Conductance It has been shown that the conductance of a tube is proportional to the cube of its radius and inversely proportional to its length Therefore it is imperative that the connecting lines be as large in dia...

Page 6: ...ze yourself with the function and action of the pump which you have now required Remove the intake and exhaust port plugs and temporarily provide a stopper for the intake and a dust cap for the exhaustReview the power requirements as described in Section 1 4 2 103 Cleanliness Take every precaution to prevent foreign from entering the pump A fine mesh screen is provided for this purpose in the inta...

Page 7: ...of the volume of the mixture this depending upon the amount of air added condensation of the vapor is wither entirely avoided or substantial reduced 2 223 Pump Function With Gas Ballast In a pump functioning on a contaminated system and operating without the gas ballast compression within the stage takes place in the normal manner until the saturation pressure of the contaminating vapor contained ...

Page 8: ...s rich in hydrogen and observing a sudden change in pressure on an electrical gauge The difference in calibration of these gauges for air and other gases will produce a distinct change in the pressure reading To use this method of detection the system must be under vacuum and the gauge sensing tube must be located between the pump and the area to be probed Use extreme caution as these materials ar...

Page 9: ... system The required frequency of changes will vary with the particular system Experiences with the process will help you determine the normal period of operation before an oil change is required 3 103 Oil Overheating This pump is designed to operate continuously below 10 Torr Continuous operation of this pump above 10 Torr will lead to overheating and eventual pump failure See Section 2 21 High P...

Page 10: ... rubber stopper Turn the pump on and run the pump for 10 minutes Close the gas ballast 8 Check the vacuum reading of the pump by connecting a thermocouple gauge tube to the pump s intake If the pump is running nearly as good as when it was new the total pressure reading you will read on wither of these two gauges will be at least 10 micron A simple way to connect the gauge tube to the pump is to r...

Page 11: ...ing surfaces cleaned of any nicks 3 50 Repairing Vacuum Leaks 3 501 Surface Preparation Good Vacuum seals are an essential and important attribute of a good mechanical pump A good seal is dependent upon the quality of the mating surfaces as well as the sealant and its preparation The mating faces should be carefully inspected for any projections or foreign particles which might interfere with prop...

Page 12: ...ut a forced oil flush 5 If noise level is still unsatisfactory consult with WELCH Technical Service at 847 676 8800 Ext 1 4 30 Excessive Oil Mist 1 Check oil level Maintain the level between the two lines on the oil window during operation Too high a level will cause excess oil mist 2 Determine the vacuum level in your system with a thermocouple gauge or manometer Be sure the pump is operating at ...

Page 13: ...02 Wired for 230V 60Hz 3Ph N A Wired for 230V 50Hz 3Ph N A Replacement Parts Shaft Seal Lip Seal 1401E 01 Shaft Seal Mechanical Seal 1401D Air Filter Screen 41 0660 Intake Chamber Cover Gasket 41 0234 Oil Case Gasket 41 0403 Drain Plug 61 2278 Oil Window 41 3753 Large Vane 41 0959 Small Vane N A Large Vane Spring 41 1304 Small Vane Spring N A Large Vane Spring Holder 41 1685 Repair Kits Minor Repa...

Page 14: ...14 Section 6 DIMENSIONAL DRAWING ...

Page 15: ...range High RPM direct drive vacuum pumps on the other hand experience a rapid fall off of pumping speed below 100 microns A comparison of the DuoSeal pumping speed curves shown below versus high RPM pump speed curves conclusively illustrates the superior performance of the DuoSeal pump What does that mean for the vacuum pump user DuoSeal pumps provide considerable faster pumpdown time in the low m...

Page 16: ...16 Section 8 EXPLODED VIEW ...

Page 17: ...or Repair Kit Catalog 1399K 03 Item 33 exploded view shown on page 18 1401E is a Mechanical Seal Kit which fits all model 1399 serial numbers There is a lipseal on pumps beginning with serial number 25773 and higher It is available as a kit part number 1401E 01 ITEM QTY P N A B DESCRIPTION 20 1 41 1239 1 1 EXHAUST VALVE 21 1 41 1285 1 1 STEEL WASHER FOR SHAFT COLLAR 22 2 41 1304 2 2 VANE SPRING 23...

Page 18: ...ST FOR PART NUMBER 41 2378 Parts List for Model 41 2378 Item Qty P N Description 1 4 2 01 0308 Screw Hex Hd 5 46 18x1 2 2 2 41 1049 Pump Leg 3 1 41 2088 Oil Case Machined 4 1 41 3753 Oil Case Window 5 1 61 2278 Drain Valve ...

Page 19: ...on kit Cat No 141601 for models1400 1405 1376 1380 1402 Standard Exhaust Filter Use with pump model Dimension Catalog Number Diameter Height Filter Replacement Element 1392 1399 1400 1376 1380 1402 1400 1397 1374 2 5 5 5 2 5 6 25 9 1417 1417P 10 1417P 20 1417L 1417G 1417H 01 Directional Exhaust Filter Use with pump model Dimension Catalog Number Diameter Height Outlet Tube O D Filter Replacement E...

Page 20: ...r Height A2 Cat No 1 NW16 to 4 CFM 140 L min 3 75 5 25 8 25 541001 Stainless Steel Element for ISO NW 16 Trap 1 541921 Stainless Steel Element for ISO NW 25 and NW 40 Traps1 541922 Removable 3 quarter center well for dry ice alcohol slurry do not use acetone Visible trapping surface Cold temperature up to 12 hours depending on vapor load Available with straight tube connections ISO Size Diameter H...

Page 21: ...uum Technology and receipt or possession does not convey any rights to loan sell or otherwise disclose said information reproduction or use of said information for any purpose other than that in connection with said information was supplied may not be made without express written permission of Welch This document is to be returned to Welch upon request and in all events upon completion of the purp...

Reviews: