Description
8
2561-67
3
Description
3.1
Design
3.1.1
Connections - Suction-/Pressure side
Suction side - Connection:
Hose nozzle DN 8
(1)
on the vacuum regulator with:
- Vacuum Manometer
(3)
- Regulating valve
(4)
- Glass Separator
(5)
for collecting liquid, which may pos-
sibly come into the pump.
The Vacuum Regulator allows for
control of the vacuum level be-
tween atmospheric pressure and
ultimate pressure of the pump.
Pressure side - Connection:
Hose nozzle DN 8
(2)
and/or Exhaust silencer
(6)
Fig. 1 Design WOB-L
®
Piston Pump (2561)
3.1.2
Connecting to the electricity supply
The WOB-L
®
Piston Pumps are supplied with complete electrical wiring. It is connected via a
mains connection cable and a power plug. Mains connection cable and plug must comply
with the requirements of the line disconnection devices (current, output).
Motor is provided with a thermal overload protection ex works, protecting the vacuum pump
from damage or destruction, respectively.
WARNING !
Should the user change the electrical connection, for example for fitting into a sys-
tem, then this may only be performed by a electrical specialist under observance of
the accident prevention regulations.
3.1.3
Display
– Vacuum Regulator
The Vacuum Regulator shows pressures at below atmospheric pressure.
The reference point is atmospheric pressure.
WARNUNG !
Please keep in mind that the barometric pressure changes from day to day.
This also changes the displayed pressure values because of latent air pressure
fluctuations.
Vacuum connection
Suction side
(1)
Exhaust connection
Pressure side
(2)
(6)
(3)
(5)
(4)