26
Step 4: Squeeze a drop of glucose control solution (Level 0, Level 1 or Level 2) onto a
clean, dry, non-absorbent surface. Do not apply control solution to the test strip directly
from the bottle. Replace the bottle cap on the control solution bottle immediately
after use.
IMPORTANT
• Do not use a test strip that has expired. Check the expiration date
XX
which is printed on the
test strip bottle.
• Use each test strip immediately after removing it from the bottle.
• After removing a test strip from the bottle, replace the bottle cap immediately and close it
tightly.
• Do not use wet or damaged test strips.
• Keep away from direct sunlight and heat
XXX
. Store the test strip bottle in a dry, cool place.
• Record the “date opened” on the test strip bottle label when you first open it. 6 months after
first opened, discard the bottle and any remaining test strips.
• Make sure you are testing in an environment between 5°C-45°C, and allow 10 to 15 minutes for it
to reach the new temperature before use. Your meter will not begin testing if it detects an out-of-
range temperature.
• You will see a flashing blood drop icon if the test strip has been inserted correctly into the test
strip port.