29
Materials you will need:
Wellion LEONARDO Cholesterol Control Solution (Level 1 or Level 2)
Your Wellion LEONARDO GLU/CHOL (plus) Meter
A new Wellion LEONARDO Cholesterol Test Strip
IMPORTANT
• Use only the Wellion LEONARDO Cholesterol Control Solution (Level 1 or Level 2) with the
Wellion LEONARDO Cholesterol Test Strip. Other brands of control solution will produce
inaccurate results.
• Always check the expiration date
XX
. DO NOT use control solutions if expired.
• Mark the newly opened bottle of control solution with the date opened. Discard any un-
used control solution 3 months after opening.
• DO NOT FREEZE or REFRIGERATE. Store the control solutions at 4°C - 30°C
Step 1: You must code the meter before performing a cholesterol control solution test
- see the section “Coding the Meter”.
Step 2: Insert a Wellion LEONARDO Cholesterol Test Strip into the meter with the blood
sample reaction zone facing up. Make sure the test strip contact points are inserted all
the way into the meter.