3
INTENDED USE
The Wellion LEONARDO GLU/CHOL (plus) Blood Glucose and Cholesterol Monitoring
System is intended to quantitatively measure blood glucose in fresh capillary whole
blood drawn from fingertips, palm, or forearm, and cholesterol in capillary whole
blood from fingertips. Alternative site testing for glucose test should be performed
only during steady-state (when glucose is not changing rapidly). Testing is done
outside the body (In Vitro diagnostic use).
The system is intended for self-testing by people with diabetes or in clinical settings by
healthcare professionals, as an aid in monitoring blood glucose level and cholesterol
level in diabetes mellitus. It is not indicated for diagnosis or screening of diabetes or
for neonatal use.
IMPORTANT:
•
Dehydration – Severe dehydration may lead to inaccurate blood glucose test re-
sults. If you suspect you are severely dehydrated, contact your healthcare profes-
sional immediately.
•
Haematocrit range – A haematocrit range that is higher than 70% or lower than 10%
can cause inaccurate blood glucose test results. A haematocrit range that is higher
than 55% or lower than 30% can cause inaccurate cholesterol test results.
•
Suitable for self-testing.
•
Not intended for use on neonates.