14
WELLION
L
UNA
DUO
Coding the Meter (Only for testing total cholesterol)
The Wellion LUNA duo meter should be coded before testing with LUNA total cho-
lesterol test strips for the first time, or every time you open a new box of LUNA total
cholesterol test strips. There are two kinds of test strips which can be used with the
Wellion LUNA duo meter; glucose test strips and total cholesterol test strips. Only total
cholesterol test strips require the code strip in order to calibrate the meter. It is not
necessary to code the meter for glucose test strips.
1. Fully insert the contact bar of the code chip into the test port of the meter, with the
labelled side of the chip facing upwards. Wait for the code number to appear on the
screen.
2. Make sure the code number and test mode appearing on the screen match the code
number and test mode marked on the test strip vial.
3. Remove the code strip by sliding down the strip ejector. “
J
” should appear on the
screen, indicating that the meter has been successfully coded for a total cholesterol test.
Cautions:
If “E-E” appears on the screen during the coding procedure, there may be a problem
with the code strip or the code strip may not have been inserted correctly. Perform
the coding procedure again, and if “E-E” continues to appear, please contact your local
distributor.