23
WELLION
L
UNA
DUO
Testing procedure for Total Cholesterol
For the consistent quality of the unused test strips, it is important to always keep them
in the original test strip vial. Use the test strip immediately after removal from the test
strip vial.
1. Remove a Wellion LUNA Cholesterol test strip from the vial and close the vial imme-
diately.
Note: Be careful not to touch the reaction zone when inserting the test strip as it may
cause an inaccurate test result.
2. Fully insert the contact bar of the test strip into the test port of the meter, with the
labelled side of the strip facing upwards. Do not bend the test strip. The Wellion LUNA
duo meter will turn on and automatically switch to the Total Cholesterol mode.
3. Check the test mode is correct - the icon “CHOL” should appear to indicate the Total
Cholesterol test mode is activated. A code number will also appear - please make sure
that this is the same as the code on the test strip vial. If the correct icon and code do
not appear, please remove the test strip and recode the meter (see chapter “Coding
the meter“). If the correct icon and code are still not displayed after re-coding the me-
ter, please remove the test strip and re-try with a new test strip. If the problem persists
with a new test strip, please contact your local distributor.