Cleaning and maintenance
06/2021
1026388-00A
Operating instructions – 63/82
4.2
Faults
Error codes are output to the display to simplify troubleshooting:
If a fault occurs, the error code appears in the display.
If several errors occur, the error messages are displayed one below the
other.
NOTE:
Some error messages can be reset by pressing the
START/STOP button.
In other cases, the fault can be eliminated by switching
off the device for 1 minute (switch-off is done at the main
switch / fuse).
Inform our customer service department if the same error code is dis-
played again after pressing the START / STOP button or switching the
device back on – see page 2.
More detailed information regarding the error description and solutions can be
found in 'Error Codes', (order no. 1020541-..).
Under certain circumstances (surge) a failure of the operating unit with touch
function may occur.
The malfunction can be eliminated by switching the unit off/on (switching
off/on of the main switch).
4.3
Cleaning and disinfection
DANGER!
Infections can be spread by skin contact!
All objects / unit parts, which may come into contact with the
user during use must be disinfected after every use:
–
Operating unit with touch function
–
Rubber blanket
–
Frame
NOTE:
Only use wipe to clean/disinfect the operating unit. Do not
spray!