Appendix
78/82
7.1.3
Table 204: Guidelines and manufacturer's
declaration – Electromagnetic interference
The wellsystem MEDWAVE TOUCH massage device has been designed for
use in the electromagnetic environment specified below. The device operator
must ensure that it is used such an environment.
Interference immunity
tests
IEC 60601 - Test level
Defined level
Electromagnetic environment
– Guidelines
Conducted HF noise IEC
61000-4-6
3 V rms 150 kHz to 80 MHz outside
the ISM band, 6 V RMS in the ISM
and amateur radio bands
Modulation:
80% AM at 1kHz
3 V rms 150 kHz to 80 MHz outside
the ISM band, 6 V RMS in the ISM
and amateur radio bands
Modulation:
80% AM at 1kHz
Portable and mobile radios are not to be
used at shorter distances to the wellsys-
tem MEDWAVE TOUCH and its cables
other than the recommended safe dis-
tance, which has been calculated ac-
cording to the transmitting frequency
are made.
Recommended safe distance:
80 MHz to 800 MHz
800 MHz to 2,7 GHz
P is the transmitter's maximum output
power (nominal rating) in Watts (W)
according to the details provided by the
manufacturer of the transmitter and the
recommended safe distance is given in
metres (m).
The field strength of the stationary radio
transmitter is for all frequencies
b
in
accordance with tests that were carried
out onsite
a
less than the determined
level.
Interference is possible in the immedi-
ate surrounding of equipment that carry
the following symbols.
Radiated HF noise IEC
61000-4-3
10 V/m 80 MHz to 2,7GHz
Modulation:
80% AM at 1kHz
27 V/m 385 MHz
Modulation:
Pulse Modulation: 18 Hz
28 V/m 450 MHz
Modulation:
FM + 5 Hz deviation: 1 kHz sine
9 V/m 710, 745, 780 MHz
Modulation:
Pulse Modulation: 217 Hz
28 V/m 810, 870, 930 MHz
Modulation:
Pulse Modulation: 18 Hz
28 V/m 1720, 1845 MHz
Modulation:
Pulse Modulation: 217 Hz
28 V/m 1970, 2450 MHz
Modulation:
Pulse Modulation: 217 Hz
9 V/m 5240, 5500, 5785 MHz
Modulation:
Pulse Modulation: 217 Hz
10 V/m 80 MHz to 2,7GHz
Modulation:
80% AM at 1kHz
27 V/m 385 MHz
Modulation:
Pulse Modulation: 18 Hz
28 V/m 450 MHz
Modulation:
FM + 5 Hz deviation: 1 kHz sine
9 V/m 710, 745, 780 MHz
Modulation:
Pulse Modulation: 217 Hz
28 V/m 810, 870, 930 MHz
Modulation:
Pulse Modulation: 18 Hz
28 V/m 1720, 1845 MHz
Modulation:
Pulse Modulation: 217 Hz
28 V/m 1970, 2450 MHz
Modulation:
Pulse Modulation: 217 Hz
9 V/m 5240, 5500, 5785 MHz
Modulation:
Pulse Modulation: 217 Hz
Comment 1: The higher value applies to 80 MHz und 800 MHz.
Comment 2: These guidelines might not apply to all situations. The propagation of electro-
magnetic waves is affected by absorption and reflexes caused by buildings, objects and peo-
ple.
a
The field strength of a stationary transmitter e.g. standard cell phone stations (mo-
bile / cordless phones) and public mobile radio services, amateur stations, MW and UHF radio
and TV transmitters) cannot be precisely determined for theoretical reasons. It is recommend-
ed that an onsite test should be carried out in order to determine the electromagnetic environ-
ment caused by stationary HF transmitters. If the field strength of the massage device deter-
mined at the installation site exceeds the defined HF level listed above then the device must
be monitored with regard to its normal operation. If any abnormal characteristics are recorded
then it will be necessary to introduce additional measures, such as reorientation or converting
the device.
b
The field strength must be less than 3 V/m over the frequency ranges from 150 kHz
to 80 MHz.