background image

3

1.  Proper Use

This wenglor product has to be used according to the following functional principle:

High-performance distance sensor

High performance distance sensors which use the principle of angle measurement determine the distance 
between the sensor and the object. These sensors have small working ranges (under 1 m) and recognize 
objects with high precision. Some sensors use a high-resolution CMOS line array and DSP signal processing. 
The color, shape and texture of the objects to be recognized does not affect the sensors’ measurements. Even 
dark objects can be reliably detected against a bright background.
They can be operated with very high speeds or very high resolutions. The measured value can be output as an 
analog value. Furthermore, Teach-In, filter functions for adjusting a switching output, and an error output are 
available. The measuring range can be selected individually within the working range.

2.  Safety Precautions

• This operating instruction is part of the product and must be kept during its entire service life. 
• Read this operating instruction carefully before using the product. 
• This product is not suitable for safety applications. 
• Installation, start-up and maintenance of this product has only to be carried out by trained personal. 
• Tampering with or modifying the product is not permissible. 
• Protect the product against contamination during start-up.

Laser/LED warning

For the respective Laser Class/LED Group please view the technical data of the product.

LASERSTRAHLUNG

NICHT IN DEN STRAHL BLICKEN

LASER KLASSE 2

EN60825-1: 2007

Pp<1.0 mW, t<2000 

µ

s, 

λ

= 620-690 nm

LASER RADIATION - 
DO NOT STARE INTO BEAM

            620 - 690 nm < 1mW

CLASS 2 LASER PRODUCT

CAUTION

Caution:

  Use of controls, adjustments or performance of procedures other than those specified herein may result  

 

in hazardous radiation exposure.

LASER CLASS 1

EN60825-1

2007

Class Laser 1 (EN 60825-1)

Observe all applicable standards 
and safety precautions.

Class Laser 2 (EN 60825-1)

Observe all applicable standards and 
safety precautions.
The enclosed laser warning labels 

 

must be attached and visible at all 
time. Do not stare into beam.

Summary of Contents for OPT2001

Page 1: ...Available as PDF only Status 28 01 2016 www wenglor com OPT2001 OPT2002 OPT2003 OPT2004 OPT2005 OPT2006 High performance distance sensor Operating Instructions EN ...

Page 2: ...cautions 3 3 Declaration of Conformity 4 4 Technical Data 5 5 Installation Instructions 8 6 Initial Operation 8 7 Function 8 8 Adjustment 9 8 1 Manual Adjustments 9 8 2 Special Settings 12 9 Maintenance Instructions 13 10 Proper Disposal 13 ...

Page 3: ...ating instruction is part of the product and must be kept during its entire service life Read this operating instruction carefully before using the product This product is not suitable for safety applications Installation start up and maintenance of this product has only to be carried out by trained personal Tampering with or modifying the product is not permissible Protect the product against con...

Page 4: ...4 EN 3 Declaration of Conformity The EC declaration of conformity can be found on our website at www wenglor com in download area RoHS ...

Page 5: ...mm Current Consumption Ub 24 V 80 mA 80 mA 80 mA Measurement Rate 1500 s 1500 s 800 s Response Time 660 µs 660 µs 1250 µs Response Time Resolution Mode 1660 µs 1660 µs 2500 µs Temperature Drift 5 µm C 10 µm C 25 µm C Temperature Range 25 50 C 25 50 C 25 50 C Analog Output 0 10 V 0 10 V 0 10 V Current Output Load Resistance 1 mA 1 mA 1 mA Analog Output 4 20 mA 4 20 mA 4 20 mA Current Output Load Re...

Page 6: ...onsumption Ub 24 V 80 mA 80 mA 80 mA Measurement Rate 1000 s 1000 s 500 s Response Time 1000 µs 1000 µs 2000 µs Response Time Resolution Mode 2000 µs 2000 µs 4000 µs Temperature Drift 5 µm C 10 µm C 25 µm C Temperature Range 25 50 C 25 50 C 25 50 C Analog Output 0 10 V 0 10 V 0 10 V Current Output Load Resistance 1 mA 1 mA 1 mA Analog Output 4 20 mA 4 20 mA 4 20 mA Current Output Load Resistance 5...

Page 7: ...ram 535 Supply Voltage V Contamination Error output NO La Emitted light disengageable O Analogue output O Ground for the analogue output Supply Voltage 0 V S Shielding Housing Dimensions Transmitter Diode Receiver Diode ...

Page 8: ...hanical regulations standards and safety precautions must be adhered to The sensor has to be protected against mechanical influences In case of very glossy surfaces the sensor has to be mounted slightly inclined and has to be mounted on a plane surface approx 5 to inhibit a direct reflection of the laser beam into the optics 6 Initial Operation Please control the proper connection of all conductor...

Page 9: ...he PNP error output is switched to positive if no object is detected within the selected working range or if an error has occurred e g an interrupted conductor at the current output The red F LED lights up to indicate that the error output has been activated 8 Adjustment 8 1 Manual Adjustments Reset Set the rotary selector switch Ý to RESET Briefly press the minus key or the plus key è red F LED l...

Page 10: ...ights up Analog output presently set to 4 20 mA Briefly press the plus key è The Yellow U LED lights up Analog output reconfigured to 0 10 V voltage output Set the rotary selector switch Ý to RUN Configuring the analog output as a current output Set the rotary selector switch Ý to U I MODE è The red F LED lights up è If the yellow U LED lights up Analog output presently set to 0 10 V è If the yell...

Page 11: ...tor switch Ý to RUN B Setting the measuring range for a rising characteristic function Zoom function Set the rotary selector switch Ý to TEACH è red F LED lights up Place the object at the most distant point of the desired working range Briefly press the plus key è yellow I LED lights up Place the object at the nearest point of the desired working range Briefly press the minus key è yellow LEDs li...

Page 12: ...N Optimize the exposure time The exposure time is adapted automatically by the sensor In the presetted DCM Default Capture Mode the Sensor has a fixed maximal possible Exposure Time It can be adjusted manually in addition in case of difficult applications By means of the LCM Long Capture Mode the Exposure Time of the sensor can be elongated for dark or highly glossy objects e g black lack in order...

Page 13: ...splay and to check the plug connections at regular intervals Do not clean with solvents or cleansers which could damage the device 10 Proper Disposal wenglor sensoric GmbH does not accept the return of unusable or irreparable products Respectively valid national waste disposal regulations apply to product disposal ...