SOP Title:
WT-
GERM KILLER™ 8910UVC BODYLight™ Operation
Effective Date:
10July2020
SOP Number:
WT-SOP-01007.05
Page:
13 of 17
Western Technology, Inc.
Confidential and Proprietary
11.5.6.
UVC BODYLight LED Operator Signals
No.:
Observed Signal:
Reason/Indicator:
Observation/Correction:
10.5.6.1.
LED
Full charge/power.
None
– Ready for Operation.
10.5.6.2.
LED
Partial charge/power.
About 1-hour of battery charge
remaining.
10.5.6.3.
LED
Low charge/power.
Recharge battery.
10.5.6.4.
Blinking
LED
Battery charge/power
too low upon
connection to light
head.
Connect a charged battery or
compliant power source.
10.5.6.5.
Flashing
Illumination
Low power signal.
Five rapid flashes indicating
about 20-minutes of charge
remaining. Replace with charged
battery.
10.5.6.6.
Blinking
LED
BODYLight has
overheated.
Thermal monitoring switched off
the BODYLight due to elevated
operating temperature. Upon
20°F drop in temperature the
BODYLight will reset and can
now be switched on.
10.5.6.7.
Repeating
-
-
LED
Blinks
UVC LED emission
power is
noncompliant to
output specifications.
BODYLight will not switch on.
Contact Western Technology for
replacement of UVC LED board.
10.5.6.8.
None
BODYLight requires
power.
Connect the BODYLight head to
a charged battery or power
source.
Ensure proper connector
contact.
10.5.6.9.
None
BODYLight does not
operate.
Contact Western Technology
Customer Service.
10.5.6.10. No Operation
BODYLight does not
operate.
Contact Western Technology
Customer Service.
10.5.6.11. Repeatedly flashing
to
LED
BODYLight has
entered UVC Code
Entry Mode.
Enter the UVC
activation/deactivation code or
press the ON/OFF button to
resume standard BODYLight
function.
10.5.6.12.
LED blinks
four times
BODYLight has
changed modes, i.e.
Visible to UVC
Correct function observed.
10.5.6.13.
LED blinks
four times
UVC activation code
accepted.
UVC mode now operative.