SOP Title:
WT-
GERM KILLER™ 8910UVC BODYLight™ Operation
Effective Date:
10July2020
SOP Number:
WT-SOP-01007.05
Page:
9 of 17
Western Technology, Inc.
Confidential and Proprietary
11.3.2.4. A charged Battery Pack can operate the UVC BODYLight for
approximately 4-hours before recharging is required.
11.4. Pre-Operation Check of the UVC BODYLight
11.4.1. Before operation of the UVC BODYLight complete a preventive
maintenance check of all components, as outlined in Section 12.2.
11.4.2. Bristle Cone Installation
11.4.2.1. Obtain the UVC BODYLight light head and the Bristle Cone.
11.4.2.2. Insert the rubber ring of the Bristle Cone into the receiving
grooves of the Bumpers on the front face of the UVC BODYLight.
11.4.2.2.1.
Note: It may be necessary to remove the top Bumper
screws in order to insert the Bristle Cone into the groove.
11.5. Operation of the UVC BODYLight
11.5.1. Put on approved PPE for potential UVC radiation exposure.
11.5.2. Attaching the Power Source to the Light Head
11.5.2.1. Firmly screw on the power source (Battery Pack, corded Battery
Pack Manifold, or corded Power Box) to the UVC BODYLight
Light Head.
11.5.2.1.1.
Ensure that a fully charged Battery Pack is used, if
applicable.
11.5.2.2. Note: When firmly hand tightened together, the gasket, located
at the base of the power source threaded connector, will
compress and a metal-to-metal contact and watertight seal will
be made.
11.5.3. Switching ON and OFF the UVC BODYLight
11.5.3.1. The UVC BODYLight is equipped with a button on each side of
the Light Head.
11.5.3.2. With the UVC BODYLight facing away from you, the Left Button
will turn ON the light and the Right Button will switch OFF the
light.
11.5.3.2.1.
Note: The Left ON Button is marked
and has the
closed/connected power
mark “ ” molded into the
fluorescent orange Bumper above this button.