5. GUIDELINES
Always allow 5 minutes for fresh liquids to degas.
Degassing is the action of removing air bubbles from the
solution, which is achieved from turning the ultrasonics on.
It is not recommended to operate unit for longer
than 30 minutes.
Fluid temperature will increase under operating conditions.
The operator should always be aware that it only takes
a small amount of moisture contamination on the power
connections to cause electrical breakdown.
•
•
•
•
•
6. DO:
Use a suitable chemical applicable to your application.
Check the suitability of the chemical for use in ultrasonics,
as some chemicals can emit a dangerous toxic gas.
Use beakers and positioning covers or auxiliary pans
for additional solutions.
Rinse thoroughly after each cleaning application.
Keep tank bottom free of silt and dirt accumulation.
Always drain the basket fully of fluid before leaving
the cleaning chamber.
✓
✓
✓
✓
✓
7. DON’T:
Run machine without solution in the tank.
Do not allow liquid level to drop below 80mm from
tank bottom.
Use heaters without a thermostat control.
Setting for the thermostat on heated models should
not exceed 60°C.
Place items directly on the tank bottom, always suspend
in basket or rack.
Drop items into stainless steel bath.
Do not pour hot water directly into the tank.
Use chemicals with very high or low pH level.
Use wax and compound removers on plastic or acrylics.
Use a flammable or combustible solution.
Subject the jacket and controls to splashing liquid
and fluid overflows.
✕
✕
✕
✕
✕
✕
✕
✕
✕
✕
✕