5.1 

Control 

The  UniSpense

®

 PRO can be used in both laboratory and industrial settings. In addition to the front-panel 

keypad, the pump can be triggered to run remotely either by the optional footswitch, or by a dry contact 
relay  or  isolated  open  collector  transistor  output  as  part  of  a  more  advanced  automated  dispensing 
system.  The    UniSpense

®

  PRO  is  capable  of  sending  an  End  of  Dispense  (EOD)  feedback  signal  to  a 

controller for remote automated operation. 

 

5.2 

User Interface 

A  chemical  resistant  membrane  keypad  allows  direct  user  entry  of  volume,  tubing  size,  automatic  delay 
interval, ramp time, pump direction, pump speed, pump function and dose count. Additional user selective 
functions include calibration, memory functions and audible cycle alert. The self-prompting, symbolic LCD 
screen ensures quick setup and ease of operation.  

 

5.3 

Performance Highlights 

An adjustable delay can be set from 0.5 to 99.9 seconds between doses to accommodate user dexterity 
and work load. The pump can be activated by depressing either the front panel mounted start key, or the 
optional footswitch for hands-free operation. 

Up  to  9  pre-selected  set-ups  can  be  stored  in  memory  and  quickly  retrieved  for  later  use.  Selections 
stored in memory are preserved even after powering the unit down. Pump speed is selectable at 75 and 
150 RPM.  A flexible tubing support stand is included to facilitate hands-free operation. 
 

6.0 

Installation 

 

Install  the  unit  where  there  will  be  adequate  room  for  the  unit  to  operate.    Provide  enough  clearance 
around the unit to keep items away from the rotating pumps and tubing.  

 

6.1 

Input Power Requirements 

This equipment is designed to  operate from a 100-240V single-phase  AC power source at  47 to 63 Hz.  
The line voltage / fuse label located on the lower rear of the unit shows the nominal input voltage set for 
the unit at the factory.    

 

6.2 

Power Cord Set 

This  unit  has  been  shipped  from  the  factory  with  a  power  line  cord  that  has  a  plug  appropriate  for  your 
area.  If  the  wrong  power  cord  has  been  shipped  for  your  particular  application,  contact  your  nearest 
WHEATON Industries dealer for the proper cord. The  UniSpense

®

 PRO has been equipped with a 3-wire 

grounding type power cord. The unit is only grounded when it is plugged into an appropriate receptacle. 

Do not operate the unit without adequate grounding protection. 

 

7.0 

Initial Startup and Operation 

 

CAUTION:

  Keep  hands  and  fingers  away  from  rotating  parts  of  the  machine.  Make  sure 

tubing is secured and clamped properly before running pump. Pump should be turned off 
and unplugged while loading tubing into pump head.  
 

 

Summary of Contents for UniSpense ELITE

Page 1: ...heatonsci com P N 50099073 Rev 2 2 10 Lit No 8734 Catalog Numbers W375040 A and W375040 B 100 120V and W375040 C through W375040 J 220 240V WHEATON UniSpense PRO Peristaltic Pump Instruction Manual In...

Page 2: ...tch Operation and Connector Functions 7 6 Flexible Tubing Holder 8 0 Unit Setup 8 1 Language Selection 8 2 Buzzer Function 8 3 Unit Reset 9 0 Preparing for Batch and Single Dose Dispensing 10 0 Making...

Page 3: ...s of identifying such a problem by calling Customer Service at 800 225 1437 U S Puerto Rico and Canada or Internationally at 856 825 1100 and then providing the Model Serial and Catalog Number of the...

Page 4: ...ain long hair 6 Wear safety goggles Wear safety goggles at all times Everyday eyeglasses only have impact resistant lenses they are NOT safety glasses 7 Don t Overreach Keep proper footing and balance...

Page 5: ...te Ranges For Various Tubing Sizes 3 2 0mm ID 0 5 mL sec and 1 0 mL sec 3 0mm ID 1 0 mL sec and 2 0 mL sec 6 0mm ID 3 5 mL sec and 7 1 mL sec 8 0mm ID 5 6 mL sec and 11 2 mL sec Vertical Head Height A...

Page 6: ...hed Retain packaging until all parts and unit functionality are confirmed Components Supplied UniSpense PRO peristaltic pump Pump Head Line Cord supplied to fit your local electrical outlet specificat...

Page 7: ...e selected set ups can be stored in memory and quickly retrieved for later use Selections stored in memory are preserved even after powering the unit down Pump speed is selectable at 75 and 150 RPM A...

Page 8: ...kinks CAUTION Improper adjustment of the tubing clamps may result in improper performance of tubing or premature wear and failure of tubing DOUBLE CHECK clamp settings before proceeding To Remove Pump...

Page 9: ...pad Functions and Icons Icons on the right of each screen show the user which keys will be active during function and setup navigation A brief explanation of active keys is listed Help Pushing the hel...

Page 10: ...or a Pump Running Pump Stopped indicator Pins 1 footswitch and 2 footswitch GND are 5V TTL compatible inputs with pull up resistors and can be driven by an isolated open collector device or dry contac...

Page 11: ...ispense alert Turning the buzzer function OFF shown will disable the buzzer during the dispense cycle 8 3 Unit Reset Press the UP and DOWN and the LEFT and RIGHT keys to access the Unit Reset Function...

Page 12: ...ate to the CALIBRATE screen 10 1 Setting Operating Speed Use the LEFT and RIGHT buttons to select the calibration speed Set the Calibration Speed so that the pump runs for at least 15 seconds during t...

Page 13: ...e Use the UP and DOWN button to navigate the Calibration Dispense icon Prepare to measure a dispense by using a graduated cylinder electronic scale or other device that can accurately measure volume P...

Page 14: ...nd DOWN and the LEFT and RIGHT keys to enter the batch count Press OK when done Entering 1 will provide a single dispense 11 2 Volume per Dispense Use the UP and DOWN keys to navigate to the Volume pe...

Page 15: ...and RIGHT keys to enter the dispense volume Press OK when done Press the Start Stop button to resume batch dispensing 11 6 To Run Additional Batches with the Same Settings Press the Start Stop button...

Page 16: ...storage of parameters If parameters are stored in a register already containing data the old data will be overwritten with the new parameters 12 2 To Recall a Parameter Press the UP and DOWN and the...

Page 17: ...n of the tubing assembly for excessive wear and leaks Many service problems can often be cured by simply replacing a worn tubing assembly To insure high accuracy the pump rotor and tubing clamping mec...

Page 18: ...pump motor speed as required Cause Motor control board defective Remedy Replace defective board Cause START Key not pressed Remedy See Unit Features and Component Identification for keypad and displa...

Page 19: ...Thickness mismatch with pump head used Remedy Tubing wall weight should match pump head specifications Inaccurate Volume Delivery Cause Tubing assembly worn Remedy Replace tubing assembly Cause Pump...

Page 20: ...IS DISCONNECTED FROM THE AC POWER SOURCE 1 Disconnect the cord set from the AC power source 2 Remove the fuse drawer by using a small flat screwdriver to lift the tab 3 Replace with new fuses 108 120...

Page 21: ...p WHEATON 800 225 1437 U S Canada Only 856 825 1100 www wheatonsci com P N 50099073 Rev 2 2 10 Lit No 8734 21 FIGURE A Major Component Identification Display Screen Keypad Pump Head I O Interface Conn...

Page 22: ...STAR 6 INT TOOTH 17 1 50099063 STF 6 32 X 1 4 ALUMINUM 18 2 WI056864 FUSE 5X20MM 2 0AT 250V 120V 18 2 WI056862 FUSE 5X20MM 1 0AT 250V FIGURES C THRU E Upper Housing Assembly Main Assembly and Wiring D...

Page 23: ...WHEATON UniSpense PRO Peristaltic Pump WHEATON 800 225 1437 U S Canada Only 856 825 1100 www wheatonsci com P N 50099073 Rev 2 2 10 Lit No 8734 23...

Page 24: ...WHEATON UniSpense PRO Peristaltic Pump WHEATON 800 225 1437 U S Canada Only 856 825 1100 www wheatonsci com P N 50099073 Rev 2 2 10 Lit No 8734 24...

Page 25: ...WHEATON UniSpense PRO Peristaltic Pump WHEATON 800 225 1437 U S Canada Only 856 825 1100 www wheatonsci com P N 50099073 Rev 2 2 10 Lit No 8734 25...

Page 26: ...WHEATON UniSpense PRO Peristaltic Pump WHEATON 800 225 1437 U S Canada Only 856 825 1100 www wheatonsci com P N 50099073 Rev 2 2 10 Lit No 8734 26...

Page 27: ...WHEATON UniSpense PRO Peristaltic Pump WHEATON 800 225 1437 U S Canada Only 856 825 1100 www wheatonsci com P N 50099073 Rev 2 2 10 Lit No 8734 27...

Page 28: ...arations of Conformity and WEEE Statement WEEE STATEMENT This UniSpense PRO peristaltic pump is not to be discarded in typical trash outlets This unit is to be discarded according to WEEE guidelines e...

Page 29: ...quirements specified by the appropriate EU Directive with respect to the design and construction of the commercialized version This declaration is invalid if modifications are performed on the device...

Reviews: